13.
Do the recommendations expire?
a.
Yes. The recommendations are only intended until the next top of the hour (e.g.,
until 4:00pm, 5:00pm, etc.). It is important to check the patient’s urine output
and other resuscitation indicators at the top of each hour. If a fluid update time
period ends within the last 15 minutes of a clock hour (e.g., from 3:45 to
3:59pm), then the software will wait until the top of the next full clock hour for
another hourly fluid update (e.g., 5:00pm).
14.
Can I use CaviCide® or similar disinfectant wipes on the screen and device?
a.
Yes, the Burn Navigator screen and device can withstand repeated use of
CaviCide wipes, Clorox Bleach® wipes, or similar disinfectant wipes. After the
CaviCide dries, there tends to be a residue visible on the screen. You can remove
the residue with a clean moist paper towel.
15.
Can I use Bluetooth to hand off a patient to my personal device?
a.
Yes, this device can exchange patient data with the Burn Nav app available on
the Google Play and Apple app stores.
16.
Can Burn Navigator connect to an electronic medical record (EMR) system?
a.
At present, no.
17.
How do I generate reports?
a.
For security reasons, this tablet cannot create or view reports, or directly
transfer report files to other devices. As a workaround, you can use Bluetooth to
hand off a patient record to another device with the app store version of Burn
Navigator. That version has the ability to generate reports. See the instructions
on the Home tab.