4
2.1
Positioning
RECOMMENDED
POSITIONS
:
favoring the
drainage of any
air present inside
the circuit.
NOT REC-
OMMENDED
POSITIONS
:
can cause the
formation of air
bubbles with
a consequent
measurement
distortion.
TANK
Ø x 10
Ø x 10
Ø
Ø
Upstream and downstream the flowmeter it is recommended to install two pipe straight lines whose length is at least 10 times the
inner diameter of OrionX (75mm x 10 = 750 mm).
Install the flowmeter far from the elements that could cause turbulence inside the pipe (valves, bends, bottlenecks, etc.).
It is recommended to position the flowmeter so that the electrodes
are not in the bottom part, in order to avoid any deposit of material
suspended in the liquid on one of the electrodes, with a consequent
measurement distortion.
ORION X must be installed at least 3 cm (1.2”) away
from metal masses that could alter the measurement.