12. After Use
The AP-Lite should always be cleaned after every use. A build-up of dirt on the inside of the
Sleeve and End Cap will cause erroneous and unstable readings.
It is advisable to clean the Probe after use with the cable attached. This will prevent
any water entering the Probe's socket and will allow any deposits to be removed
from the connector collar and shell.
The Sleeve on the AP-Lite can be removed by
unscrewing to allow cleaning of the electrode.
After every use, remove the Sleeve End Cap then
unscrew the Sleeve.
Rinse the exposed electrode, the inside of the Sleeve and
the Sleeve End Cap with fresh, clean water.
Dry the Electrode and polish the lenses using a soft cloth.
Shake the water from inside the Sleeve and Sleeve End
Cap, then reattach.
Dry the outside of the Probe using a soft cloth.
TIP:
Occasional application of a smear of silicone grease to the connector O-rings and
thread, Sleeve thread, the Sleeve End Cap thread will make fitting and removal of these
parts easier.
Never clean the Probe with solvents, alcohol or concentrated acid/alkaline based
cleaning products such as Decon 90. These products can strip the anodised finish
from the Probe and damage the plastic and rubber components. Damage caused by
the use of aggressive cleaning agents or solvents is not covered by your warranty.