Aquaprobe
®
Lite Instruction Manual
10401-00976 Rev F
SELECT ELECTRODE
1:Oil
l
4:EMPTY
2:EMPTY
l
5:EMPTY
3:EMPTY
l
6:EMPTY
The Oil electrode should have been assigned to an AUX socket when it was fitted. Choose
that socket. Press the OK or right arrow key to select Oil. The screen will change to:
CALIBRATE Oil
ZERO? [01/Jan/12]
Pt-2? [01/Jan/12]
Calibration point 2 (Pt-2) is the upper calibration point.
The dates shown to the right of each point are the dates of the last successful calibration.
7.
Select ZERO. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Output:2500mV
Calibrating
100%
Press [OK]
The Calibration Report on the top line displays the voltage output from the electrode in
millivolts (mV). This value is not stored in memory so should be noted down in a calibration
record book for the probe.
11.9.8.
Calibrating Point 2
Remove the Probe from the calibration cup, shake off any excess water then dry the outer
sleeve with a soft cloth.
Pour 300mL of freshly mixed 1-5, naphthalenedisulfonic acid disodium salt calibration
solution into a clean calibration cup then gently lower the Probe in all the way.