background image

AQUACLAVE Manual Flushing Membrane System

RO Membrane

Cartridge Filters

04

Cartridge Filters

Filter Description

Service Life

Stage 1
Pretreatment

This cartridge commences the treatment

process by removing suspended solids
such as dirt and rust. The other important

function is to reduce the chlorine and

organic content of the feed water

6 Months

Stage 2
Pretreatment

The purpose of this high performance

carbon filter is to further remove free
chlorine and organic content of the water.
It completes the 2 stage pretreatment of

the feed water supply.

6 Months

Stage 3
RO TFC Membrane

This specifically engineered low volume

production capacity with high rejection

TFC membrane is manufactured in the

USA. The membrane removes
approximately 99.99% of all dissolved
minerals and salts present in the feed

water supply. Production and performance
will be influenced by the chemical analysis

of the water supply.

3 Years

Stage 4
Post Carbon Filter
Organic Terminator

NSF approved. This post membrane
carbon filter is designed to remove
organic residue. It removes residual
impurities and odors from the tank and
provides a finer conditioning of pure

water.

6 Months

Stage 5
Non-organic Polishing
Cartridge

This final process addresses all non-

organic residue leaving a quality of water

that complies with 

AS 2182 1998

Appendix C Quality of Feed Water for

Steam  Generation 

when the system is

operated in accordance with this manual.

6 Months

Carbon Filter

Carbon Filter

Post Carbon Filter

Non-organic Filter

Summary of Contents for AquaClave P5

Page 1: ...t 15 Memo USER S MANUAL Aqualife Products Australia Pty Ltd Steam Sterilisor Water System Thank you very much for selecting AQUALIFE In order to bring the best use of your system please read the user s manual carefully before installation and follow the regulations Attention This is a scientific instrument Please ensure that only Aqualife approved replacement components are used Inferior prefiltra...

Page 2: ... Drinking water solutions are also available from Aqualife Products Australia Pty Ltd IMPORTANT The system has been engineered to operate within a controlled environment Where high water pressure risks exist above 860 Kpa 125 psi a 600 Kpa pressure limiting valve must be installed to avoid nulliying the warranty Australian Plumbing Industry regulations require the installation of a dualcheck valve...

Page 3: ... The total dissolved solids TDS test meter should be used on every ocasion when drawing water for use Specifications Production 50LPD 240Volt 50 60Hz with a transformer Safety approval TUV EN60950 CE MARK The system has been designed to comply with Australian electrical standards Special technical designs Patented No 068018 flush membrane 100 factory tested and sterilized ready for installation Op...

Page 4: ...on system the goal is not to dilute the salt solution but to separate the pure water from the dissolved salts and other contaminants When the natural osmotic flow is reversed water from the salt solution is forced to pass through the membrane in the opposite direction by application of pressure thus the term REVERSE OSMOSIS Through this process we are able to produce pure water by screening out th...

Page 5: ...AQUACLAVE Manual Flushing Membrane System Components selected filters 03 Components 11 litres water storage tank Filter Housing wrench Delivery valve Water supply Tee connector Waste Drain saddle ...

Page 6: ...h high rejection TFC membrane is manufactured in the USA The membrane removes approximately 99 99 of all dissolved minerals and salts present in the feed water supply Production and performance will be influenced by the chemical analysis of the water supply 3 Years Stage 4 Post Carbon Filter Organic Terminator NSF approved This post membrane carbon filter is designed to remove organic residue It r...

Page 7: ...AQUACLAVE Manual Flushing Membrane System 05 The parts of AquaClave RO P5 5th To faucet 5th Stage to tank ...

Page 8: ...connection diagram CONNECTING THE COLOR TUBING A Connect the WHITE tubing to the water supply Tee connector B Connect the BLUE tubing to the sink top faucet C Connect the BLACK tubing to the waste drain saddle D Connect the RED tubing to the storage tank ...

Page 9: ...AQUACLAVE Manual Flushing Membrane System Installation diagram 07 AQUALIFE MEMBRANE T connector RO membrane Install tundish before drain saddle above trap ...

Page 10: ...AQUACLAVE Manual Flushing Membrane System 08 Change filters 5 Litres Turn Open housing clockwise with the wrench flush ...

Page 11: ...AQUACLAVE Manual Flushing Membrane System 09 Change membrane Turn using pliers Loosen Loosen the Jaco fitting anticlockwise Jaco ...

Page 12: ...AQUACLAVE Manual Flushing Membrane System Operation regulation 10 Turn ...

Page 13: ... FOR LEAKS E After initial flushing open the shut off valve on the tank and close the sink top faucet F Tank will now be full of water usually 5 to 6 hours Open the sink top faucet and drain all water until the storage tank is empty and there is only a small flow from the sink top faucet DO NOT USE FIRST TANK OF WATER G Close the sink top faucet The system is now ready for use H Change filters reg...

Page 14: ...hing function We suggest you apply the manual flushing function when A This is the first time you start to use the AquaClave system B When it is time to change filters Applying the manual flushing function regularly once a week is also recommended Caution Keep the system running when manual flushing function is on Have the manual flushing function on for 30 seconds Do remember to turn the manual f...

Page 15: ...f R O water to fall to half of that at 23 C The maximum water temperature recommended is 27 C 3 Total Dissolved Solids TDS The higher the amount of dissolved contaminants in the water the lower the quantity of water produced A high level of TOTAL DISSOLVED SOLIDS can be overcome with additional water pressure 4 Membrane Different membranes have different characteristics Some produce more water tha...

Page 16: ...AQUACLAVE Manual Flushing Membrane System Maintenance checking list 14 ...

Page 17: ...AQUACLAVE Manual Flushing Membrane System Memo 15 AquaClave RO P5 ...

Reviews: