3.4.2 SDI12 frame.
A list of SDI12 registers is available for network communication.
Please ask to our hotline service the documentation on SDI12 communication.
4. Installation.
4.1 Sensor installation option
For the installation of the sensors in condition of immersion or in-pipe insertion, we advise to use
accessories adapted and proposed by AQUALABO.
4.1.1 Accessories for immersion installation.
In immersion condition, it is necessary to maintain the sensor by the body and not to leave the sensor
suspended by the cable at the risk of damaging the sensor.
AQUALABO proposes a range or pole (long version) in order to install the sensor in open basins. It
can be positioned a considerable distance from the basin edge with the bracket suspended for
example.
Please note the following when planning your set-up:
• The fitting must be easily accessible to allow the sensor or the fitting itself to be maintained
and cleaned regularly
• Do not allow the fitting (and thus also the sensor) to swing against and hit the basin edge
• When working with systems involving pressure and/or temperature, ensure that the fitting
and sensor meet all relevant requirements
• The system designer must check that the materials in the fitting and sensor are suitable for
the measurement (chemical compatibility, for instance)
Material
PVC
Admissible temperature
0 to 60 °C
Pressure max.
5 bars