10
Version 1.0
Update: August 2018
AQUALABO
90 Rue du professeur P. Milliez
94506 Champigny-sur-Marne, FRANCE
Tel.: +33(0)1 55 09 10 10
4.1.2 Accessories for PVC pipe-mounting.
Every system of assembly is delivered with an adapter (and the appropriate joints) and one T of
assembly (90 ° for CTZN sensor) to stick on a 90 mm diameter pipe. Its special design type ensures
the correct inflow to the sensor, thus preventing incorrect measurements.
Please note the following when planning your piping set-up:
• The fitting must be easily accessible to allow the sensor or the fitting itself to be maintained
and cleaned regularly
• We recommend bypass measurements. It must be possible to remove the sensor through
the use of shut-off valves
• When working with systems involving pressure and/or temperature, ensure that the fitting
and sensor meet all relevant requirements
• The system designer must check that the materials in the fitting and sensor are suitable for
the measurement (chemical compatibility, for instance)
Mounting system for CTZN sensor (
PF-ACC-M-00001
)
1 Unscrew the union nut (2) from the PVC flow fitting (3).
2 Guide the sensor cable through the union nut on the fitting.
3 Insert the sensor (1) into the fitting as far as the position shown in the middle image above.
4 Screw the union nut onto the fitting as far as the stop.
1
3
4
CTZN sensor
Union nut
Adapter
90 mm pipe diameter
2