3. Description. 

 

3.1 Product overview 

 

Inductive method : 

A  ring-type  coil  is  excited  at  fixed  intervals  and  the  response  is  retrieved  on  a  second  coil,  which  is 
linked to the excited coil. The connectivity between the coils (determined by the degree of conductivity) 
takes place via the conducting solution. 

Economic and successful technology that requiring not enough maintenance and not consumable. 

The CTZN sensor offers the following advantages: 
• Low operating costs due to reduced maintenance work (no electrolyte changes) 
• Greater calibration intervals due to low drift behaviour 
•  No  polarization  voltage  required  •  High  measuring  accuracy,  even  for  low  concentrations  •  Rapid 
response times 
The  sensor  features  excellent  interference  immunity  thanks  to  the  integrated  preamplifier  and  digital 
signal  processing.  The  measured  value  for  conductivity  is  automatically  compensated  with  the 
temperature  and  transferred  without  interference  to  the  connected  display  unit  and  controller  via  a 
digital interface. The sensor also includes a log book containing the last ten successful calibrations in 
the form of a ring buffer. 

 

3.2 Applications 

 

The compact and robust sensor is particularly well suited to the following typical areas of application: 
• Urban wastewater treatment  
• Industrial effluent treatment 
• Surface water monitoring 
• Sea water 
• Fish farming 
 
3.3 Construction and dimensions. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

(1) NTC Temperature sensor 
(2)  EPDM Head of body  
(3)  PVC Sensor body with measurement electronics 
(4) Cable bushing 
(5) Securely connected connection cable 

 

 

Summary of Contents for Ponsel CTZN

Page 1: ...1 Version 1 0 Update August 2018 AQUALABO 90 Rue du professeur P Milliez 94506 Champigny sur Marne FRANCE Tel 33 0 1 55 09 10 10 CTZN NUMERICAL SENSOR User manual ...

Page 2: ...3 1 Product overview 6 3 2 Applications 6 3 3 Construction and dimensions 6 3 4 Communication 7 3 4 1 Modbus RTU registers 7 3 4 2 SDI12 frame 8 4 INSTALLATION 8 4 1 Sensor installation option 8 4 1 1 Accessories for immersion installation 8 4 1 2 Accessories for PVC pipe mounting 10 4 1 3 Accessories for stainless steel pipe mounting 11 4 3 Electrical connections 12 5 STARTUP AND MAINTENANCE 13 5...

Page 3: ...iliar with and comply with the instructions in this manual Make sure the power supply complies with the specifications before connecting the device A clearly labeled power switch must be installed near the device Check all connections before turning the power on Do not attempt to use damaged equipment it may represent a hazard and should be labeled as faulty Repairs must only be carried out by the...

Page 4: ...ge temperature 10 C à 60 C Maximum refreshing time Maximum 1 seconde Sensor Dimensions Diameter max 62 4 mm Lenght 196 mm Weight 700 g Wetted Material Body EPDM PVC Stainless steel Cable polyurethane jacket Steam gland Polyamide Maximum pressure 5 bars Connection 9 armoured connectors polyurethane jacket bare wires or waterproof Fisher connector Protection IP68 Sensor cable Standard 3 7 and 15 m o...

Page 5: ...liance with the European standards Standard tests EN 61326 1 edition 2013 Emission EMC EN 55022 Class B Immunity EN 61000 4 3 A EN 61000 4 2 B EN 61000 4 6 A EN 61000 4 4 B Shone disturbances EN 55011B Identification of the measurement process composed of 1 one probe 2 Ponsel s cable EN 61000 4 5 Not concerned for sensors with a cable lower or equal to 30 M Commercial name DIGISENS range Manufactu...

Page 6: ...e required High measuring accuracy even for low concentrations Rapid response times The sensor features excellent interference immunity thanks to the integrated preamplifier and digital signal processing The measured value for conductivity is automatically compensated with the temperature and transferred without interference to the connected display unit and controller via a digital interface The ...

Page 7: ...ameter temperature of the Sensor and to calibrate the parameter temperature Furthermore there are certain numbers of functions such as Select the averaging value Read the Sensor description Return to default coefficients Modify the Sensor address Information on measures conducted Out Of Specification measures measures in progress etc Date and name of the operator who performed the calibration etc ...

Page 8: ...e cable at the risk of damaging the sensor AQUALABO proposes a range or pole long version in order to install the sensor in open basins It can be positioned a considerable distance from the basin edge with the bracket suspended for example Please note the following when planning your set up The fitting must be easily accessible to allow the sensor or the fitting itself to be maintained and cleaned...

Page 9: ...le and specially studied to adapt themselves to the different configurations of assembly Elbowed pole with elbowed shutter PF ACC C 00109 90 ELBOW LONG POLE FOR CTZN SENSOR 2955 mm ELBOWED SHUTTER Straight long pole with elbowed shutter PF ACC C 00115 STRAIGHT LONG POLE FOR CTZN SENSOR 2745 mm ELBOWED SHUTTER Elbowed shutter Vertical axis Connection vertical pole horizontal axis Connection horizon...

Page 10: ...ined and cleaned regularly We recommend bypass measurements It must be possible to remove the sensor through the use of shut off valves When working with systems involving pressure and or temperature ensure that the fitting and sensor meet all relevant requirements The system designer must check that the materials in the fitting and sensor are suitable for the measurement chemical compatibility fo...

Page 11: ...e acceptable maximum pressure for the sensors is 5 bars The system of assembly can be delivered with or without stainless steel clamp The adapter is compatible with a 90 mm diameter external clamp Mounting system for CTZN sensor PF ACC C 00350 1 After welding the clamp 3 on the stainless steel pipe remove the clamp from the system and remove the PVC adapter 1 2 Guide the sensor cable through the u...

Page 12: ...3 2 V Consumption Standby 25 µA Average RS485 1 measure seconde 6 3 mA Average SDI12 1 measure seconde 9 2 mA Current pulse 500 mA Protection against the inversions of polarity Wiring diagram Cable length up to 15m 1 Red Power supply V 2 Blue SDI 12 3 Black Power supply V 4 Green B RS 485 5 White A RS 485 6 Green yellow Cable shield with Power supply V Cable lenght 15 to 100 meters Red Purple Yell...

Page 13: ...lization before measure processing Started 5 2 Calibration The calibration of the conductivity sensor is a 2 step process Step 1 offset Expose the sensor to the air to perform the first stage of the Calibration process The value for this first calibration standard is set to 0 0 μS cm Step 2 gain the sensor is placed in a buffer solution of known conductivity 12 88 mS cm Let stabilize during 5 minu...

Reviews: