AquaCare: Installation
A. Ensure the unit is turned off and is fully depressurised. If the machine has been ON
recently, allow 30 seconds for depressurising before attempting to remove the lid.
B. Unscrew the required dosing chamber lid anti-clockwise 2/3 of a turn to relieve the
pressure. Then unscrew a further turn and lift off.
C. Select the required powder, i.e. Prophy or Prep, and shake the required Powder
Cartridge for 10 to 20 seconds (See TREATMENTS, Pg 26-30).
D. Pull out top yellow insert from the cartridge.
E. Push in the clear lower draw tab on the cartridge.
F. Line up the rib on the outside of the powder cartridge with the slot on the inside of
the required dosing chamber - Prophy (Left) or Prep (Right).
G. Slide the powder cartridge down into the dosing chamber until it sits comfortably
inside. If the cartridge sits unevenly or above the top of the dosing chamber, then lift
it out and re-insert. Check that the rib of the powder cartridge slides down the slot in
the dosing chamber - refit the dosing chamber lid.
H. Close the dosing chamber by turning lid clockwise. Turn firmly until a strong resistance
is felt from the seal engaging with the top of the chamber. Continue to turn the lid
until you feel the lid tighten against a stop.
Always make sure the lid is fully closed
before pressurising the machine.
I. To remove a powder cartridge, open the dosing chamber lid and gently lift the
cartridge out of the machine. If the cartridge still contains some powder and will be
used again, carefully pull back to lock the lower tag in position, DO NOT REMOVE IT,
then refit the top insert.
All powder cartridges supplied by Velopex are gamma irradiated. The contents are sterile
until opened. Powder cartridges must not be refilled as the cartridges are not reusable.