Assembling the handpiece
AquaCare
Fitting the Disposable Feedline and Tip
A. Take a new tip and fold over the locking tab as shown.
B. Push the tip onto the nozzle until firmly engaged.
C. Rotate the tip so that the free end points towards the handpiece.
D. Push one end of the feedline onto the end of the handpiece.
E. Push the other end of the feedline onto the tip.
The tips and feedlines are for one time use only, a new tip and feedline should be used for
each patient. To switch tips and feedlines remove the existing feedline tip and discard. Then
follow the steps above. Purge the fluid through the system.
See Maintenance: Purging the AquaCare Pg 33.
WARNING: The handpiece should be sterilised before use and after each
subsequent use. See Maintenance: Sterilising the Handpiece, Pg 31.
Feedline
A. See Parts and Features: Handpiece, Pg 12. Slide the handpiece handle onto the twin tube.
This goes at the end of the tube without a check valve.
B. Push the twin tubes firmly onto the nozzle tubes. The small twin tube pushes onto the
small nozzle tube, and the large twin tube pushes onto the large nozzle tube. Make sure
the twin tubes are securely in place.
C. Slide the handle onto the handpiece until it is securely in place.
D. Push the free end of the larger twin tube into the ‘Powder Out’ connector on the machine,
see page 10 item 2.
E. Push the free end of the smaller twin tube onto the ‘Fluid Out’ connector on the machine,
see page 10 item 3.
F. See page 42 for further information.