EN
1
Sterile filtration: Path-
ogen retention ≥ 10
7
B.
diminuta
CFU/cm
2
filter
membrane surface
(= 7 log steps).
2
Measured in accordance
with the international
ASTM F838-15a stand-
ard – Standard Test
Method for Determining
Bacterial Retention of
Membrane Filters Uti-
lized for Liquid Filtration.
3
Depending on water qual-
ity. If you use the Aqua
free
filter change service,
the filter will be changed
every 4 or 8 weeks.
4
The Baclyser
®
neo
T
must not be connected
to low pressure units
(boiler).
5
Continuous dosage of
≤ 10 ppm over operating
time; short-term (1h) high
dosage (400,000 ppm)
for chemical disinfection.
Technical data
Size:
49.3 mm x Ø 52 mm
Retention:
7 log steps
2
Membrane:
Flat sheet membrane,
pore size 0.2 μm
Operating time:
up to 31, 62 or 93 days
(1M, 2M or 3M)
3
Flow rate:
15.5 l/min at 5 bar,
13.1 l/min at 3 bar
Max. operating pressure:
5 bar
4
Max. operating temperature:
60 °C
(at 75 °C ≤ 90 min over operating time)
Connection:
Quick-release coupling
Chlorine resistance:
≤ 10 ppm
5
Chlorine dioxide resistance:
≤ 0,8 mg/l
49
.3 m
m
52 mm