Running PCR Samples 4-7
About the Run Time
Screen
You can use the Run Time screen to chart progress at any time during the run. The
Run Time screen displays the executing segment, and the next segment to execute.
5
Press
Start
to start a run.
If the temperature of the heated cover is less than 103 °C, this message, “Cover is
heating” appears.
6
When the heated cover reaches 103 °C, the
Run Time
screen displays and the
method you selected starts running.
For a description of this screen, see “About the Run Time Screen” on page 4-7.
To start a run:
(continued)
Step
Action
Current temperature: 65°C
The run will begin when the
heated cover reaches 103°C.
Please wait. Cover is heating
F1
F3
F4
F5
F2
Cancel
Sample 56.7
Cycle 1 of 25
HOT
55.0
94.0
2:00
10:00
55.0
0:30
72.0
0:30
72.0
5:00
4.0
94.0
0:30
F1
F3
F4
F5
F2
Pause
Info
Use the Run Time screen for...
For more information, see page
Stopping a Run Before It Completes
Pre-PCR segment
PCR segment
Post-PCR segment
Calculated sample temperature
updated as run progresses
HOT displays when
the sample block
exceeds 50 °C
Temperature
parameters are
displayed above the
graph
Horizontal lines flash as
hold times decrement
Cycles count up when
run is in PCR segment
55.0
94.0
2:00
10:00
55.0
0:30
72.0
0:30
72.0
5:00
4.0
94.0
0:30
F1
F3
F4
F5
F2
Pause
Info
Sample 56.7
Cycle 1 of 25
HOT
Ramps flash when system is
approaching next temperature value
Hold times are displayed
below the graph
Summary of Contents for GeneAmp PCR System 9700
Page 1: ...GeneAmp PCR System 9700 Base Module User s Manual ...
Page 9: ...ix To Obtain Customer Training Information C 7 D Limited Warranty Statement Index ...
Page 10: ......
Page 28: ......
Page 56: ......
Page 90: ......
Page 96: ......
Page 110: ......
Page 116: ......
Page 124: ......
Page 126: ......
Page 131: ......