29
[SETUP] Page
Figure 5-3 Stray Admittance
[OPEN CORR] function will correct all frequencies. The frequency points
may be different in different versions.
Procedure of open correction
Step 1
Press [SETUP] to enter <SETUP> page
Step 2
Keep test clip or cable open circuit.
Step 3
Press [OPEN CORR] soft key
Step 4
Press [OK] to execute.
When correction is done, the data will be saved.
During correction, press [Cancel]to cancel correction.
5.2.2
Short Correction
Short correction compensates any remaining impedance.
USB Command: CORR:SHORT
Figure 5-4 Remaining impedance
Procedure of short correction
Step 1
Press [SETUP] key to enter <SETUP> page
Step 2
Use the short jumper to shorten test clips or connect cables
together.
Step 3
Press [SHORT CORR]
Step 4
Press [OK] key to start. During correction press[Cancel]to
cancel correction.