7. ELIMINATING DEFECTS
Should any anomalies occur during use, it is necessary to contact the owner’s qualified
personnel trained for the maintenance and repair of the bed.
The user shall in no case try to eliminate any malfunctions that might occur in the electrical
equipment!
Dangerous for your life, life threat, due to electric shock!
Any interventions to the electrical equipment must be carried out only by
qualified service and repair staff members, by the producer or by qualified and
licensed electricians, in accordance with VDE and relevant safety standards!
Possible solutions
The following table is useful to eliminate possible problems and faults:
Problem
Possible causes
Solution
The hand control and /
or the operating system
does not work
·
Power cord not connected
·
Plug not under voltage
·
Hand control plug not properly
connected
·
Faulty parts
·
Insert the power cord
·
Check the socket / the fuse box
·
Check all connections
·
Inform the owner and ask for repairs
Pressing the keys, the
movements are
activated only for a
short time
·
Too much weight on the bed
·
The bed hits against an obstacle
·
Reduce the load
·
Remove the obstacle
Movement in one
direction only
·
Defective hand control or
operating console
·
Inform the owner and ask for repairs
Function in contrast
with the instruction on
the hand control
·
Reversed motor connectors
·
Inform the Manufacture and ask for
repairs
The movements stop
immediately after a
longer adjustment time
·
Activation of the circuit breaker in
the transformer
·
Put in a rest position and let the system
cool down for about 20 to 30 minutes