21
Accessories
Fixing, possible damages
Inspector’s signature:
Control result:
Date:
·
If you think that there might be damage or a functional problem, the bed must not
be used at all and it must be disconnected from the power supply until the damaged
components have been totally replaced or repaired!
·
To replace or repair damaged parts, please contact the owner of the bed.
6.2 By the owner
The owner of this nursing bed is required to carry out regular tests in case of any new
installation, any maintenance service and during the normal use operations to ensure the
safety of this bed.
These tests must be verified during the regular maintenance services depending on use
conditions.
·
The functional tests, the evaluations, the documentation of the inspection results must be
carried out only by experienced personnel who know the proper way and conditions
necessary to implement a correct testing.
·
The evaluation and the documentation of the results can be performed only by skilled
electricians with additional medical and technical knowledge.
·
The controls must be carried out according to the sequence below:
1.
Visual inspections
2.
Electrical measurements
3.
Functional tests
Frequency of the inspections
As a benchmark, we recommend an examination every six months can be extended in
accordance with the terms of use on its own responsibility respecting a share of proven error
of 2% to a maximum of two years.
·
If you think that there might be a damage or a functional problem, the bed
must not be used at all and it must be immediately disconnected from the power
supply until the damaged components have been totally replaced
or repaired!