8. SERVICE 

 
The VARILYMPH 12 PRO is equipped with high quality components and is subject to strict quality requirements. 
Please get in contact with the SLK GmbH in case of any defects.  
SLK VARILYMPH 12 PRO devices can the maintained and repaired at SLK GmbH.  
To ensure the safety of the system over a long period, it must be performed at least every 2 years a safety inspection 
(STK) to the system. This can be done at the company SLK or an authorized dealer. 
If there are any questions please don´t hesitate to contact the SLK GmbH. 
 
8.1 Error messages and troubleshooting 
The  VARILYMPH 12  PRO  is programmed in a way that the function of the device is monitored by electronics during 
operation. The following error messages may occur 
 
 
Warning "Check for leakage": the electronics has determined that the selected target pressure could not be established. 
The reason for this may be a leak in cuffs, hoses and hose connectors. Please check the connections of the hoses to the 
device; please pay particular attention that the connectors are fully engaged on all four edges. The white clips on the 
connector must not protrude after correct locking: 
 

 

 
 
If the message does not disappear after checking of connectors, then please check also the other parts of the hose system 
for leakages (watch out for hissing noises). If you can find no cause, so please contact the responsible dealer 
 
Warning "Error 10": Here the electronics has determined that a movable component that is responsible for the filling of 
the treatment sleeves, could not go to the start position. Reason for this could be storage or operation at temperatures 
below 15 ° C. In this case please increase the ambient temperature. If the device has been stored at low temperatures 
(eg, overnight in a car) so please wait until the device has adapted to the ambient temperature. If the warning persists, 
you should contact the responsible dealer. 
 
Should incidents other than those described here appear, so please contact the responsible dealer before operation. 
 

 
 
 
 

 
 
 
 
 
 

Summary of Contents for VARILYMPH 12 PRO

Page 1: ... 06 2019 USER MANUAL VARILYMPH 12 PRO CONTENTS 1 ABOUT 2 SAFETY INSTRUCTIONS 3 SCOPE OF SUPPLY 4 FIELD OF APPLICATION 5 OPERATION INSTRUCTION 6 END OF THERAPY 7 CLEANING 8 SERVICE 9 TECHNICAL SPECIFICATIONS ...

Page 2: ... 2 5 The operation should be done only in closed rooms and by trained personnel or medical personnel 2 6 Danger when using the system in the presence of easy flammable gasses 2 7 Keep the control unit from water and moisture eg steam away If water has penetrated into the control unit disconnect it immediately from the power supply 2 8 Please do not operate the control unit before it warmed up or c...

Page 3: ... the therapy with IPC ask your attending doctor 5 OPERATION INSTRUCTIONS To put the device into operation connect the power cable with the power supply and put the plug into the power outlet Connect the device and the garments with the connection tubes Make sure that the grey plugs of the garments are plugged into the grey socket of the device Make sure that the cover cap closes the second tube co...

Page 4: ...irm by pressing the deflation time button again inflation time Press the inflation time button Adjust the setting by turning the dial and confirm by pressing the inflation time button again You can also wait for approx 8 seconds to confirm the settings The display will switch to the start menu automatically target pressure Adjust the target pressure in the start menu directly turn the dial clockwi...

Page 5: ... and remove organic deposits by using a neutral detergent and a water moistened cloth Avoid eccess of water or cleaning solution on the surface of the compressor The SLK pressure cuffs can also be cleaned with a water moistened cloth If heavily soiled the pressure cuffs can be washed also in the washing machine at a low spin up to max 60 C Caution ensure that the connection hoses to the pressure c...

Page 6: ... hoses and hose connectors Please check the connections of the hoses to the device please pay particular attention that the connectors are fully engaged on all four edges The white clips on the connector must not protrude after correct locking If the message does not disappear after checking of connectors then please check also the other parts of the hose system for leakages watch out for hissing ...

Page 7: ... operation Temperature 15 C 40 C Max rel humidity 85 not condensing Presssure 700mbar to 1060mbar Transport and Storage Temperature 20 C 50 C Max rel humidity 85 not condensing Presssure 700mbar to 1060mbar medical device class IIa protection class II insulation protected device For VARILYMPH 12 PRO a manufacturer s warranty of 2 years applies All parts and data are in constant development and may...

Page 8: ...V2 7 06 2019 SLK Vertriebsgesellschaft mbH Am Herdicksbach 18 45731 Waltrop GERMANY phone 49 0 231 925360 0 fax 49 0 231 925360 29 email info slk gmbh de web www slk gmbh de ...

Reviews: