6. END OF THERAPY
The device turns off automatically when the treatment time is up. The device changes over to standby mode. Disconnect
the tubes to deflate the garments. Always wait till the garments are fully deflated before opening the zipper to get out of
the garments.
For easier transport of the system, please loosen the hose connections to the device. This avoids damage to the control
unit and the pressure cuffs.
Quick - Evacuation
The therapy can be terminated at any time by disconnecting the unit from the mains. Treatment cuffs deflate automatically
during a power failure.
7. Cleaning and disinfection
Please note the attached cleaning instructions for SLK systems.
Cleaning
Clean all exposed surfaces and remove organic deposits by using a neutral detergent and a water-moistened cloth. Avoid
eccess of water or cleaning solution on the surface of the compressor.
The SLK pressure cuffs can also be cleaned with a water-moistened cloth. If heavily soiled, the pressure cuffs can be
washed also in the washing machine at a low spin up to max. 60°C.
Caution: ensure that the connection hoses to the pressure cuff a securely closed with the enclosed plugs before laundry.
They prevent the ingress of water.
Chemical disinfectants
We recommend disinfection with disinfectants based on alcohol (70%).
Moisten all surfaces with disinfection solution, then wipe with a water-moistened cloth and then dry it thoroughly.
For detailed information on cleaning and reprocessing in the SLK-Hygiene Center, please contact your dealer or directly
the company SLK.
The fuse can be changes in the little compartment between power switch and E14 socket: