VARILYMPH 12 PRO

 

 

 

 

CONTENTS

 

1.  ABOUT 
2.  SAFETY INSTRUCTIONS 
3.  SCOPE OF SUPPLY 
4.  FIELD OF APPLICATION 
5.  OPERATION INSTRUCTION 
6.  END OF THERAPY 
7.  CLEANING 
8.  SERVICE 
9.  TECHNICAL SPECIFICATIONS 

 
 

 

 

 

 

 
 
 
 
 
 

 
 
 
 

 

Summary of Contents for VARILYMPH 12 PRO

Page 1: ... 06 2019 USER MANUAL VARILYMPH 12 PRO CONTENTS 1 ABOUT 2 SAFETY INSTRUCTIONS 3 SCOPE OF SUPPLY 4 FIELD OF APPLICATION 5 OPERATION INSTRUCTION 6 END OF THERAPY 7 CLEANING 8 SERVICE 9 TECHNICAL SPECIFICATIONS ...

Page 2: ... 2 5 The operation should be done only in closed rooms and by trained personnel or medical personnel 2 6 Danger when using the system in the presence of easy flammable gasses 2 7 Keep the control unit from water and moisture eg steam away If water has penetrated into the control unit disconnect it immediately from the power supply 2 8 Please do not operate the control unit before it warmed up or c...

Page 3: ... the therapy with IPC ask your attending doctor 5 OPERATION INSTRUCTIONS To put the device into operation connect the power cable with the power supply and put the plug into the power outlet Connect the device and the garments with the connection tubes Make sure that the grey plugs of the garments are plugged into the grey socket of the device Make sure that the cover cap closes the second tube co...

Page 4: ...irm by pressing the deflation time button again inflation time Press the inflation time button Adjust the setting by turning the dial and confirm by pressing the inflation time button again You can also wait for approx 8 seconds to confirm the settings The display will switch to the start menu automatically target pressure Adjust the target pressure in the start menu directly turn the dial clockwi...

Page 5: ... and remove organic deposits by using a neutral detergent and a water moistened cloth Avoid eccess of water or cleaning solution on the surface of the compressor The SLK pressure cuffs can also be cleaned with a water moistened cloth If heavily soiled the pressure cuffs can be washed also in the washing machine at a low spin up to max 60 C Caution ensure that the connection hoses to the pressure c...

Page 6: ... hoses and hose connectors Please check the connections of the hoses to the device please pay particular attention that the connectors are fully engaged on all four edges The white clips on the connector must not protrude after correct locking If the message does not disappear after checking of connectors then please check also the other parts of the hose system for leakages watch out for hissing ...

Page 7: ... operation Temperature 15 C 40 C Max rel humidity 85 not condensing Presssure 700mbar to 1060mbar Transport and Storage Temperature 20 C 50 C Max rel humidity 85 not condensing Presssure 700mbar to 1060mbar medical device class IIa protection class II insulation protected device For VARILYMPH 12 PRO a manufacturer s warranty of 2 years applies All parts and data are in constant development and may...

Page 8: ...V2 7 06 2019 SLK Vertriebsgesellschaft mbH Am Herdicksbach 18 45731 Waltrop GERMANY phone 49 0 231 925360 0 fax 49 0 231 925360 29 email info slk gmbh de web www slk gmbh de ...