20
Navigation and Customization
How to Tare
First, ensure that nothing is flowing through
the device.
Taring Flow
MENU
→
TARES
→
TARE FLOW
. Flow tares should
occur at the expected process pressure, as long
as there is no flow. A message, "
ENSURE NO
FLOW BEFORE PRESSING TARE
" will be displayed.
Press
TARE
to complete the taring process.
Taring Pressure
MENU
→
TARES
→
TARE PRESS
. Absolute pres-
sure tares must be done with the meter open to
atmosphere (requires a barometer, which is included by default in portable
devices). A message, "
PRESS TARE WHEN VENTED TO AMBIENT WITH NO FLOW
"
followed by "
CURRENT PRESSURE OFFSET
:" If the absolute pressure sensor
in the flow body disagrees with the internal barometer, the difference will
be displayed. If there is no difference, it will display:
----- PSIA
.
When to Tare
•
Before every new flow measurement cycle
•
After significant changes in temperature or pressure
•
After dropping or bumping the flow meter.
•
After installing the meter in a different orientation.
Menu
You can enter the menu system by pressing
the
MENU
button from the Main Display.
ABOUT TARES
BASIC
ADV
CONFIG SETUP
MAIN
Main Menu
•
ABOUT
),
TARES
(below),
BASIC CONFIG
), and
ADV SETUP
(
) enter
their respective menus.
•
MAIN
exits to the Main
Display (
Taring Your Flow Meter
Taring is an important practice that ensures
that your flow meter is providing the most
accurate measurements possible. This function gives the flow meter a
good zero reference for flow measurements. For meters with an optional
barometer, taring can also be used to align the internal absolute pressure
sensor with the barometric pressure reading.
TARE
TARE
PRESS
FLOW
BACK
MAIN
Tare Menu