ORP60-Z Smart ORP/Redox Tester 

(

ORP/Temp.)

 

Instruction Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

APERA INSTRUMENTS, LLC 

www.aperainst.com 

V1.03

 

 

Summary of Contents for ORP60-Z

Page 1: ...1 SKU AI3717 ORP60 Z Smart ORP Redox Tester ORP Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03...

Page 2: ...arranty 8 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leaves factory It does NOT mean...

Page 3: ...dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following gra...

Page 4: ...n off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibrat...

Page 5: ...ed e g distilled deionized water for a long time which will make the electrode response slow If this happens soak the ORP electrode in 3M KCl solution for 3 5 hours and then re calibrate it before usi...

Page 6: ...ound 20 C then adjust the calibration value to 227 mV Notes 1 Usually there s no need to calibrate for ORP testing When the tester has been used for a long time users can test it in 222 mV ORP standar...

Page 7: ...For example if 10 seconds is set when the measured value is stable for more than 10 seconds the measured value will be automatically locked and the HOLD icon will be displayed Short press to release...

Page 8: ...ery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our...