- 6 - 

 

ORP Calibration

     

 

7.1 Pour some 222 mV ORP standard solution into the calibration bottle (to about half volume). 

 

7.2 Rinse electrode in distilled water and dry it. Short press            to turn the meter on. Long press            to enter 

calibration mode. If you need to exit calibration mode, short press 

 

7.3 Dip electrode in the standard solution, stir gently, and allow it to stand still in the calibration solution until a stable 

reading is reached. When stable icon            comes up and stays, press            to adjust the calibration value according to 

the table below(also on thesolution bottle

’s label). Then short press            to complete the calibration. 

 

222 mV (25˚C) ORP Standard Buffer Solution Calibration Reference Table 

Temperature (˚C) 

mV 

Temperature (˚C) 

mV 

10 

242 

30 

215 

15 

235 

35 

209 

20 

227 

38 

205 

25 

222 

40 

201 

 

For example, if the temperature is around 25˚C, then adjust the calibration value to 222 mV. If the temperature is around 

20˚C, then adjust the calibration value to 227 mV.   

 

Notes 

1) 

Usually there’s no need to calibrate for ORP testing. When the tester has been used for a long time, users can 

test it in 222 mV ORP standard buffer. If the error is big, please calibrate according to the steps in 7.1 to 7.3. 

2)  The meter has self-diagnosis function. Please refer to the followed chart for detailed information. 

Symbol 

Self-Diagnosis information 

How to fix 

 

 

 

              is pressed before the 

measurement becomes stable 

(smiley face comes up) 

Wait till the smile icon appears and stays, then short press                           

 

During calibration, readings being 

unstable for over 3 minutes 

1. 

Please check whether ORP electrode is damaged or broken. 
If so, please replace with a new one. 

2. 

The platinum pin or junction is severely contaminated. 
Please use a soft brush with soap water to clean it 
thoroughly. Then soak it in 3M KCL overnight before 
performing calibration again. 

3. 

The electrode is aged (used for over a year and has a much 
slower response). A replacement is needed. 

If none of the above, please contact us at 

[email protected]

 or +1 

(614) 285-3080 

Summary of Contents for ORP60-Z

Page 1: ...1 SKU AI3717 ORP60 Z Smart ORP Redox Tester ORP Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03...

Page 2: ...arranty 8 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leaves factory It does NOT mean...

Page 3: ...dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following gra...

Page 4: ...n off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibrat...

Page 5: ...ed e g distilled deionized water for a long time which will make the electrode response slow If this happens soak the ORP electrode in 3M KCl solution for 3 5 hours and then re calibrate it before usi...

Page 6: ...ound 20 C then adjust the calibration value to 227 mV Notes 1 Usually there s no need to calibrate for ORP testing When the tester has been used for a long time users can test it in 222 mV ORP standar...

Page 7: ...For example if 10 seconds is set when the measured value is stable for more than 10 seconds the measured value will be automatically locked and the HOLD icon will be displayed Short press to release...

Page 8: ...ery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our...

Reviews: