background image

 

- 5 - 

 

  Complet e Kit  

 

 

 

 

 

 

 

 

 

Things to Know  Bef ore Use  

1) 

A few drops of water are added to the  probe cap to keep the pH electrode in an activated state  before the tester 

leaves factory. Generally, users can start using the tester directly. If the measuring response is slow, users can soak 

the ORP electrode for 30 minutes in the 3M KCL storage solution (fill to the fill-line) before using it; if the electrode is 

dry for a long time (> 1 month), the ORP electrode responds slowly, You can soak the ORP electrode for 8 hours in 

the storage solution before using. 

2) 

When the tester is not in use, we recommend adding one to two drops of tap water to the probe cap, and close the 

cap (be careful not to add too much water). This way, the ORP probe’s sensitivity can be maintained and users can 

start using the tester right away next time. 

3) 

Do not soak the ORP electrode in purified (e.g. distilled/deionized) water for a long time, which will make the electrode 

response slow. If this happens, soak the ORP electrode in 3M KCl solution for 3~5 hours, and then re-calibrate it 

before using. 

4) 

The storage solution is 3M KCL (SKU: AI1107), and the tester kit comes with a bottle of 10mL storage solution (can 

be used repeatedly). If it is contaminated, replace it with a new one. Please do not use other brands’ storage solutions 

as they may contain other chemicals that can cause damage to the electrodes. 

5) 

Things needed in addition to what’s in the box: 

a) 

Distilled or deionized water (8-16oz) for rinsing the probe after each test 

b) 

Tissue paper for drying the probe 

 

ORP M easurement  

ORP stands for Oxidation-Reduction Potential, measured in mV. 

It’s also called redox. ORP is a measure of the 

cleanliness of water & its ability to break down contaminants. Rinse the probe in distilled water and dry it. Dip the probe 

in sample solution, shake for a few seconds, and allow it to stand still. Get the ORP readings after            appears and 

stays on screen. 

Graph-2 

 

Summary of Contents for ORP60-Z

Page 1: ...1 SKU AI3717 ORP60 Z Smart ORP Redox Tester ORP Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03...

Page 2: ...arranty 8 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leaves factory It does NOT mean...

Page 3: ...dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following gra...

Page 4: ...n off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibrat...

Page 5: ...ed e g distilled deionized water for a long time which will make the electrode response slow If this happens soak the ORP electrode in 3M KCl solution for 3 5 hours and then re calibrate it before usi...

Page 6: ...ound 20 C then adjust the calibration value to 227 mV Notes 1 Usually there s no need to calibrate for ORP testing When the tester has been used for a long time users can test it in 222 mV ORP standar...

Page 7: ...For example if 10 seconds is set when the measured value is stable for more than 10 seconds the measured value will be automatically locked and the HOLD icon will be displayed Short press to release...

Page 8: ...ery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our...

Reviews: