222 mV (25˚C) ORP Standard Buffer Solution Calibration Reference Table 

Temperature (˚C) 

mV 

Temperature (˚C) 

mV 

10 

242 

30 

215 

15 

235 

35 

209 

20 

227 

38 

205 

25 

222 

40 

201 

 
For example, if the temperature is around 25˚C, then adjust the calibration value to 222 mV. If the 
temperature is around 20˚C, then adjust the calibration value to 227 mV.  

press 

  to complete the calibration. 

 

6. 

ORP Measurement 

Press 

 to power on, clean the probe in distilled water and dry it. And place the probe in the 

tested solution and stir gently and let it stand still. Read the measurement after       comes up and 
stays. 

 

7. 

Notes

 

1) 

Usually there’s no need to calibrate for ORP testing. When used for a long time, users can 
test it in 222 mV ORP standard buffer. If the error is big, please calibrate according to the 
steps in 5.1 to 5.4. 

2)  The meter has self-diagnosis function. Please refer to the followed chart for detailed 

information. 

 

Symbol 

Self-Diagnosis information 

How to fix 

 

 

 

       is pressed before the 
measurement becomes stable 
(smiley face comes up) 

Wait  till  the  smile  icon  appears  and 
stays, then short press              

 

7.1 

The cleaning and activation of ORP probes 

After long-time use, the surface of the platinum sensor can be contaminated and cause 
inaccuracy and slow response time. If that happened, the followed methods can be adopted to 
clean and activate: 

a)  For contamination caused by inorganic substances, dip the probe into 0.1 mol/L dilute 

hydrochloric acid for 30 minutes. Clean it with distilled water, soak it into the KCL 
storage solution for 6 hours, and then it will be ready for use again. 

b)  For contamination caused by organic substances and oiliness, clean the platinum 

surface with dishwashing liquid, and wash with distilled water. And then the probe will be 
ready for use again after 6 hours’ soaking in the storage solution. 

c)  If the platinum surface is so severely contaminated that there forms an oxide film, 

toothpaste can be used to clean the platinum surface, and then wash it with distilled 
water. The probe will be ready for use again after 6 hours’ soaking in the storage 
solution. 

 
 

 

Summary of Contents for 28APE190027

Page 1: ...ORP60 Premium Pocket ORP Tester APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...t 1 In calibration mode press to change calibration value unidirectional 2 In mode setting short press to change parameter unidirectional 1 Long press to enter calibration mode 2 In calibration mode s...

Page 3: ...on 1 Clean electrode in distilled water and dry it 2 Pour certain amount of 222 mV ORP standard buffer solution into the calibration bottles about half volume of the bottle 3 Short press to turn the m...

Page 4: ...How to fix is pressed before the measurement becomes stable smiley face comes up Wait till the smile icon appears and stays then short press 7 1 The cleaning and activation of ORP probes After long ti...

Page 5: ...of 120 mV and 150 mV Preset low value 120 mV highest value 150 mV when the measured value 120mV or 150mV stable icon displays on LCD Red back lit heads up displays B Automatic lock P3 Select On to ac...

Page 6: ...table measurement Reading value auto hold HOLD Self Diagnostic Information Er2 Low Voltage warning flickering reminder of battery replacement Three color backlight Blue Measurement Mode Green Calibrat...

Page 7: ...ARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Transportation storage improper use failure to follow the product instructions or to perform any...

Page 8: ...APERA INSTRUMENTS Europe GmbH Wilhelm Muthmann Stra e 15 42329 Wuppertal Germany Contact info aperainst de Website www aperainst de Tel 49 202 51988998...

Reviews: