GENERAL
162/184
NTA47271-300A2
CALYS 1500
You may apply scaling to a device: select "ON" and set parameters.
Press key F2 (Points) to set calibration points.
Use the following buttons to define the points.
to edit an existing calibration point,
to add a calibration point,
to delete a calibration point.
Use the F1 (
) button to select the transmission mode field, press the function button, F4 (
) to drop down
the menu and use the Up
/ Down
arrow buttons to choose the transmission mode.
Definitions of transmission modes:
Manual: the settings are generated under manual control with the user accepting each point,
One-Way: the settings are generated automatically in the order in which the points (Point 1, Point 2, Point3 etc.) are
defined,
Two-Way: the settings are generated automatically in the order in which the points (Point 1, Point 2, Point3 etc.) are
defined and then in reverse order (Point n, Point (n-1) etc.).