Calibrated sensors
C
NTA47271-300A2
153/184
C.1.6
Cyclic Ramp Mode
The cyclic ramp generation function is used to program a first linear variation in a direction (increasing or decreasing)
followed by a first step and then a second linear variation in a direction opposite to the first variation followed by a second
step.
When this mode is active, the
icon appears in the transmission window.
The
function key is used to launch an increasing cyclic ramp and the
function key is used to launch an
decreasing cyclic ramp.
The default parameters of this mode are displayed in the left hand side of the transmission window. To change these
parameters, press ENTER or use the Menu
Mode….
The parameters of a cyclic ramp are:
B: minimum amplitude of the signal.
H: maximum amplitude of the signal.
Thb: duration of a decreasing ramp.
Tbh: duration of an increasing ramp.
Th: duration of the high step.
Tb: duration of the low step.
Nbr: number of cycles to be generated.
Delay: delay in seconds between launch of the cyclic ramp and the start of transmission.