GENERAL
146/184
NTA47271-300A2
CALYS 1500
The following ranges are available:
Range
10000 Hz
100000 Hz
Resolution
0.01 Hz
0.1 Hz
Max. Amplitude
20 V
20 V
Depending on the range selected, display is in Hz or beats per minute (BPM).
Connection of the frequency source generated is between the Hz and COM terminals.
B.3.1.9
Pulse generation
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Pulse transmission function, then the range using the function and navigation keys.
Select the “Type of Output” Signal.
Enter the amplitude of the signal between 0 and 20 V.
Confirm with ENTER.
The
icon appears in the Transmission window.
To change the default parameters, press ENTER or enter the Configuration … menu then Pulses ….