For Veterinary use only
Switch ON the suction pump and Install the biopsy valve; immerse the distal end of the endoscope into
the detergent and depress the suction valve to aspirate detergent into the instrument channel for 30s;
remove the distal end from the detergent and cover the suction valve to aspirate air for 10s.
Switch OFF the pump and disconnect the endoscope from the endoscope imaging processor; remove
suction valve and biopsy valve from the endoscope.
Attach the water resistant cap onto the endoscope connector; put the endoscope, air/water valve
(inapplicable to VET-3512), suction valve and biopsy valve into container and transport them to the
cleaning and disinfection room.