04
Operation
- 13 -
For Veterinary use only
Turn the UP/DOWN and RIGHT/LEFT angulation locks to the
“F ” direction to release them.
Turn the UP/DOWN and RIGHT/LEFT control knobs to their respective neutral positions; confirm that the
bending section returns to an approximately straight condition
Carefully withdrawal the endoscope from the body while observing the endoscopic image.
4.4 Video Endoscope Withdrawal
CAUTION
⚫
The high surface temperature of the front end of the endoscope connector may be caused by the
transmission of high energy light. Do not contact it immediately with skin and wait for at least 2
minutes for safe contact.
⚫
The power switch of the endoscope imaging processor must be shut OFF, and then disconnect
the videoscope cable before disconnecting the endoscope from the imaging processor.
4.5 Cleaning and disinfection of the endoscope
After each examination, disconnect the endoscope from the processor and clean it, especially
when the distal end and the instrument channel has contacted with mucus. Insufficient cleaning
and disinfection time may lead to cross-contamination between patients and undesired
performance of the endoscope.
After each clinical examination, cleaning and disinfection should be conducted immediately.
The cleaning and disinfection method is determined by the user and hospital infection control
committee, etc.
The cleaning and disinfection include automatic and manual method. AOHUA only introduces
recommendation on manual cleaning and disinfection methods. For the automatic method, refer
to the instruction manual for relevant equipment.
1
2
3