File Management
4-4 File Menu
Site Master User Guide
PN: 10580-00252 Rev. J
4-11
Save On Event Menu
Key Sequence:
File
>
Save On Event
In Spectrum Analyzer mode, this menu is used to auto save measurements
to internal memory after:
...Crossing Limit:
Toggling this submenu key to On will save the
measurement to internal memory when the measurement has crossed a
defined limit line created with the
Limit
menu.
...Sweep Complete:
Toggling this submenu key to On will save the
measurement to internal memory after the current sweep is complete. If
Save Then Stop is toggled Off, a measurement will be saved after every
sweep.
Save Then Stop:
Set this key to On to stop the sweep after a
measurement is saved. With this key Off and Sweep Complete On a
measurement is saved after every sweep.
Clear All:
Pressing this key will turn Off the three save on event keys:
Crossing Limit
Sweep Complete
Save Then Stop
Figure 4-11.
Save On Menu
Back
Clear All
Save On...
...Crossing Limit
On Off
...Sweep Complete
On Off
Save Then Stop
On Off