Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-43
FMS <value> <unit>
Description:
Opens the external FM sensitivity parameter.
Functional
Category:
PARAMETER ENTRY
Value:
±10 kHz/V to ±20 MHz/V in Locked, Locked Low-Noise, and Unlocked Narrow FM;
±100 kHz/V to ±100 MHz/V in Unlocked Wide FM
Units:
KV, MV
Example:
FML; FMS 500 KH
Turns on FM external locked with a sensitivity of 500 kHz/Volt.
Output Command:
OFS
(in Hz)
Related
Commands:
FM1
,
FM2
,
FML
,
FMU
,
FMW
Front Panel Key:
Modulation
|
FM
(external) |
Edit Sensitivity
FMU
Description:
Turns on the external FM function in unlocked narrow mode (Same as FM1).
The signal generator output is not phase-locked. Disables the internal FM function or the
external FM function in Unlocked Wide, Locked, or Locked Low-Noise mode, had any of
these modes been previously programmed.
Functional
Category:
MODULATION
Related
Commands:
FM0
,
FM1
,
FM2
,
FML
,
FMW
Front Panel Key:
Modulation
|
FM
(external) |
Mode > | Unlocked Narrow
then
Modulation
|
FM
(external) |
On
FMW
Description:
Turns on the external FM function in unlocked wide mode.
The signal generator output is not phase-locked. Disables the internal FM function or the
external FM function in Unlocked Narrow, Locked, or Locked Low-Noise mode, had any
of these modes been previously programmed.
Functional
Category:
MODULATION
Related
Commands:
FM0
,
FM1
,
FM2
,
FML
,
FMU
Front Panel Key:
Modulation
|
FM
(external) |
Mode >
|
Unlocked Wide
then
Modulation
|
FM
(external) |
On
Summary of Contents for MG369 C Series
Page 2: ......
Page 6: ...Contents 4 PN 10370 10374 Rev F MG369xC GPIB PM ...
Page 262: ...A 34 PN 10370 10374 Rev F MG369xC GPIB PM ...
Page 264: ...Index 2 PN 10370 10374 Rev F MG369xC GPIB PM ...
Page 265: ......