Chapter 3 — Power Sensor Care
3-5
RF Connector Care
Connectors should never be forced together when making a
connection since forcing often indicates incorrectness and
incompatibility. There are some dimensions that are critical for
the mechanical integrity, non-destructive mating and electrical
performance of the connector. Connector gauge kits are available
for many connector types. Please refer to Anritsu Application
Note 10200-00040. The mechanical gauging of coaxial connectors
will detect and prevent the following problems:
Positive Pin Depth
Positive pin depth can result in buckling of the fingers of the
female center conductor or damage to the internal structure of a
device due to the axial forces generated.
Negative Pin Depth
Negative pin depth can result in poor return loss, possibly
unreliable connections, and could even cause breakdown under
peak power conditions
.
Checking the Pin Depth Gauge
Pin depth gauges should be checked for cleanliness before they
are used at a minimum of once per month. Connector cleaning
procedures (refer to
) can also be used to clean the pin
depth gauges.
Caution
Never make a connection when any positive pin depth
condition exists.