Post Calibration
Actions
After the calibration process is completed, perform the following ac-
tions, as appropriate:
If the calibration passes:
Save the calibration data to (hard) disk, as follows:
Step 1.
Press the
Option Menu
key (Enhancement key group) to dis-
play the OPTIONS menu.
Step 2.
Select DIAGNOSTICS from menu, then select in se-
quence: H/W CALIBRATIONS | DISK OPERATIONS |
SAVE TO HARD DISK | HW_CAL.LO2.
If the calibration fails:
Step 1.
Repeat the calibration.
Step 2.
If the calibration fails, replace the A2 PCB and perform the
calibration again.
LO2 CALIBRATION
ADJUSTMENTS
6-6
37XXXD MM