Operator’s Manual
Rev 1/24/20
pg. 33
Controller Temperature Calibration
1.
Fill the instrument with 2 liters of tap water.
2.
Press the red
HEAT
and blue
AGITATE
buttons on the front of the
instrument so that they are lit.
3.
Bring the water to a boil.
4.
Press
on the controller until you see
A
Ac
.
Cd
l
.
5.
Press
on the controller until you see
A
Ac
.
04
l
on Display 2.
6.
Press
on the controller until you see
A
CAL
.
H
l
on Display 2.
7.
Stir the water vigorously.
8.
Place a calibrated thermometer in the bottom of the Vessel near the probe without the thermometer touching the
Vessel. This will be your reference temperature.
9.
Press
on the controller to set Display 2 to the temperature on the calibrated thermometer.
10.
Press
on the controller until you see
A
Ac
.
Cd
l
on Display 2.
11.
Press
on the controller until you see
A
Ac
.
01
l
on Display 2.
12.
Press
on the controller to save the calibration.
The ANKOM
200
controller is now calibrated. The instrument will boil water at the calibrated temperature and maintain
the proper temperature throughout an assay.
Display 2
Display 1
tab up down enter
Summary of Contents for ANKOM200 FIBER ANALYZER
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 6: ...Operator s Manual pg 6 Rev 1 24 20 This page intentionally left blank ...
Page 12: ...Operator s Manual pg 12 Rev 1 24 20 This page intentionally left blank ...
Page 18: ...Operator s Manual pg 18 Rev 1 24 20 This page intentionally left blank ...
Page 28: ...Operator s Manual pg 28 Rev 1 24 20 This page intentionally left blank ...
Page 36: ...Operator s Manual pg 36 Rev 1 24 20 Appendix A Wiring Diagram ...
Page 37: ...Operator s Manual Rev 1 24 20 pg 37 This page intentionally left blank ...