Page 75
Alerts, Warnings and Alarms
14
Warning: Exceeds Max Basal
Cause
Basal delivery (or Temp Basal) exceeds user-set
maximum.
Effect
Basal delivery stops.
Message
Displayed when awakened (by basal delivery attempt
every 3 min. or manually) until confirmed.
Action
Press OK/Enter to confirm. Reprogram Max Basal
amount in Setup Advanced (or reprogram Temp Basal).
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Delivery Canceled Due to Low Cartridge
Cause
Basal or Bolus delivery exceeds insulin remaining in
cartridge.
Effect
Basal or Bolus delivery stopped.
Message
Once per occurrence and each time awakened until
confirmed.
Action
Press OK/Enter to confirm. Replace with full cartridge.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: No Prime, No Delivery
Cause
Pump is not primed.
Effect
All deliveries stop.
Message
Every 3 minutes or when awakened manually.
Action
Press OK/Enter to confirm. Disconnect, reprime.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Alarm: Occlusion
Cause Occlusion
detected.
Effect
All deliveries stop.
Message Continuous
until
confirmed.
Action
Press OK/Enter to confirm. Disconnect and prime to clear
occlusion. Option to select Suspend (see Suspend).
Beeps/Vib
User-selected every 3 minutes. If not confirmed,
progresses to sweep/vibe within one hour. (Once
confirmed, No Prime alarm triggered, see No Prime.)
Summary of Contents for IR 1200
Page 1: ...Page 1...
Page 6: ...Page 6...
Page 11: ...Page 11 Radiology Quick Reference Chart 1...
Page 16: ...Page 16...
Page 18: ...Page 18...
Page 22: ...Page 22...
Page 32: ...Page 32...
Page 50: ...Page 50...
Page 56: ...Page 56...
Page 68: ...Page 68...
Page 70: ...Page 70...
Page 82: ...Page 82...
Page 99: ...Page 99...