Page 74
14
Alerts, Warnings and Alarms
Warning: Low Cartridge
Cause
Low insulin level reached.
Effect
Deliveries may continue until Empty Cartridge alarm is
triggered.
Message
Displayed when manually awakened until confirmed.
Action
Press OK/Enter to confirm. Replace with filled cartridge.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Exceeds Max Bolus
Cause
Audio bolus delivery exceeds user-set maximum.
Effect
Bolus delivery stops.
Message
Displayed when manually awakened until confirmed.
Action
Press OK/Enter to confirm. Reprogram max bolus
amount in Setup Advanced.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Exceeds Max TDD
Cause
Bolus delivery exceeds user-set maximum.
Effect
All deliveries may stop until action is taken.
Message
Displayed when manually awakened until confirmed.
Action
Press OK/Enter to confirm. Reprogram max TDD amount
in Setup Advanced.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Low Battery
Cause
Battery life ends in minimum 1 hour.
Effect Deliveries
continue.
Message
Displays when pump is awake until confirmed. Displays
when triggered by event (such as bolus) & when
manually awakened.
Action
Press OK/Enter to confirm. Insert new battery.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Summary of Contents for IR 1200
Page 1: ...Page 1...
Page 6: ...Page 6...
Page 11: ...Page 11 Radiology Quick Reference Chart 1...
Page 16: ...Page 16...
Page 18: ...Page 18...
Page 22: ...Page 22...
Page 32: ...Page 32...
Page 50: ...Page 50...
Page 56: ...Page 56...
Page 68: ...Page 68...
Page 70: ...Page 70...
Page 82: ...Page 82...
Page 99: ...Page 99...