17
1,Connect the microscope to computer via the USB cable included. After the microscope monitor
displays few options, choose “USB camera”by pressing button “down”&”OK” or using remote.
2
,Open software“Microscope Measure”.
3,Click “File”→ “Open”→ “Open Device”→ “USB MODE” → “USB CAMERA”.
*
Functions
1
)
Calibration
1, Put the ruler under the digital microscope, adjust the focusing tube and the height of the
stand, to get the best clearance and desired magnification. (After its done, keep the object
distance for the rest steps)
2, In the 'Calibration management' set the name and unit length of the new calibration(refer
the picture on the right). Then, click 'Calibration', meanwhile, the '
' in the toolbar should
have been chosen automatically. If no, please choose it yourselves.
3, Move the cursor to the main operation window, draw a unit length line with help of the
ruler(it doesn’t matter if the length result shows differently). After these, click the 'Finish'
button in the bottom of the 'Calibration Management' window.
4, Check. The length of the line you draw in step 3 should now appear as the unit length. The
name of the new calibration should have showed in the list of calibrations.