background image

 

DN:KD-7961N-SMSY01 V3.0 

Operation Guide 

Template version:1

 

10/15 

12. TECHNICAL ALARM DESCRIPTION 

The monitor will show 'SYS HI','SYS Lo' ,'DIA HI' or 'DIA Lo' as technical alarm on LCD with no delay if the 

determined blood pressure (systolic or diastolic) is outside the rated range specified in part SPECIFICACIONS. 

In this case, you should consult a physician or check if your operation violated the instructions. 

The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or 

inactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8. 

The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear automatically 

after about 8 seconds.

 

13.TROUBLESHOOTING (1) 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

LCD 

Display 

shows 

abnormal result 

The  cuff  position  was  not  correct  or  it 

was not properly tightened 

Apply the cuff correctly and try again 

Body  posture  was  not  correct  during 

testing 

Review 

the 

―BODY 

POSTURE 

DURING 

MEASUREMENT‖  sections  of  the  instructions  and 

re-test. 

Speaking,  arm  or  body  movement, 

angry, excited or nervous during testing 

Re-test  when calm and  without speaking  or moving 

during the test 

Irregular heartbeat (arrhythmia) 

It is inappropriate for people with serious arrhythmia 

to use this Electronic Sphygmomanometer. 

 

14. TROUBLESHOOTING (2) 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

LCD Display shows battery 

symbol

 

Low Battery 

Change the batteries 

LCD Display shows 

 

―EE‖ 

Arm or blood pressure monitor was 

moved during testing 

Re-test taking care to not move your arm or 

the blood pressure monitor 

The cuff does not inflate properly or 

pressure falls quickly during testing 

Make certain the plugs and brackets of the 

cuff are securely attached to the monitor. 

The cuff was not properly applied. 

Apply the cuff correctly and try again 

No response when you 

press button or load battery. 

Incorrect 

operation 

or 

strong 

electromagnetic interference. 

Take out batteries for five minutes, and then 

reinstall all batteries. 

In Clock Mode, no response 

when you press button. 

The touch switch is in OFF position. 

Please keep touch switch in ON position, 

and try again 

 

Summary of Contents for KD-7961N

Page 1: ...TERY LOADING 5 2 TOUCH BUTTON SWITCH 6 3 CLOCK AND DATE ADJUSTMENT 6 4 VOICE SETTING 6 5 CONNECTING THE CUFF TO THE MONITOR 7 6 APPLYING THE CUFF 7 7 BODY POSTURE DURING MEASUREMENT 7 8 TAKING YOUR BLOOD PRESSURE READING 8 9 DISPLAYING STORED RESULTS 8 10 DELETING MEASUREMENTS FROM THE MEMORY 9 11 ASSESSING HIGH BLOOD PRESSURE FOR ADULTS 9 12 TECHNICALALARM DESCRIPTION 10 13 TROUBLESHOOTING 1 10 1...

Page 2: ...ht Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and active Considering the above information it is recommended that you measure your blood pressure at approximately the same time each day Too frequent measurements may cause injury due to blood flow interference please always relax a minimum moment of 1 to 1 5 minute...

Page 3: ...1 General requirements for basic safety and essential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnetic compatibility Requirements and tests IEC 80601 2 30 2009 Cor 2010 EN 80601 2 30 2010 Medical electrical equipment Part 2 30 Particular requirements for ...

Page 4: ...cuff should be placed at the same level as your heart 4 During measurement neither speak nor move your body and arm 5 Measuring on same wrist for each measurement 6 Please always relax a minimum moment of 1 to 1 5 minutes between measurements to allow the blood circulation in your arm to recover Prolonged over inflation cuff pressure exceed 300 mmHg or maintained above15 mmHg for longer than 3 min...

Page 5: ...umidity ranges in specifications 15 Please do not share the cuff with other infective person to avoid cross infection 16 This blood pressure monitor is verified by auscultatory method It is recommended that you check annex B of ANSI AAMI SP 10 2002 A1 2003 A2 2006 for details of verification method if you need 17 This equipment has been tested and found to comply with the limits for a Class B digi...

Page 6: ...nd LCD will display time and date See picture 3 Picture 3 Picture 3 1 Picture 3 2 Picture 3 3 Picture 3 4 b While the monitor is in Clock Mode pressing both the START and MEM button simultaneously touch button switch in ON position the month will blink at first See picture 3 1 Press the button START repeatedly the day hour and minute will blink in turn See picture 3 2 3 3 3 4 While the number is b...

Page 7: ...he monitor until the plugs and brackets are securely attached 6 APPLYING THE CUFF a Place the cuff around a bare wrist 1 2cm above the wrist joint on the palm side of the wrist b While seated place the arm with the cuffed wrist in front of your body on a desk or table with the palm up If the cuff is correctly placed you can read the LCD display c The cuff must be neither too tight nor too loose No...

Page 8: ...pressure has built up for a measurement Then the monitor slowly releases air from the cuff and carries out the measurement Finally the blood pressure and pulse rate will be calculated and displayed on the LCD See picture 8 3 The irregular heartbeat symbol if any and blood pressure classification indicator will blink on the screen The result will be automatically stored in the memory If the voice f...

Page 9: ...pressing button MEM for three seconds all results in the memory bank will be deleted after three beep See picture 10 picture 10 1 Press the button MEM or START the monitor will turn off Picture 10 Picture 10 1 11 ASSESSING HIGH BLOOD PRESSURE FOR ADULTS The following guideline for assessing high blood pressure without regard to age or gender has been established by the World Health Organization WH...

Page 10: ...ng Review the BODY POSTURE DURING MEASUREMENT sections of the instructions and re test Speaking arm or body movement angry excited or nervous during testing Re test when calm and without speaking or moving during the test Irregular heartbeat arrhythmia It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer 14 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION...

Page 11: ...priately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or three years and the cuff integrity is maintained after 1 000 open close cycles of the closure 10 It is recommended the cuff should be disinfected 2 times every week if ...

Page 12: ...magnetic emissions The KD 7961N is intended for use in the electromagnetic environment specified below The customer or the user of the KD 7961N should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The KD 7961N uses RF energy only for its internal function Therefore its RF emissions are very low and are not...

Page 13: ...terial the relative humidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Guidance and manufacturer s declaration electromagnetic immunity The KD 7961N...

Page 14: ...magnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters a...

Page 15: ...W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the transmitter wh...

Reviews: