  DN:KD-322-SMSY01 V4.0 

Operation Guide 

Template version:1

 

4/14 

 

SPECIFICATIONS 

 

1.  Product name: Blood Pressure Monitor 

2.  Model: KD-322 

3.  Classification:  Internally powered, Type  BF applied part,IPX0,No  AP  or APG,Continuous 

operation 

4.  Machine size: approx. 144mm×102mm×46mm (5 21/32

” x 4 1/32” x 1 13/16”) 

5.  Cuff  circumference:  22cm-30cm(8  21/32

” -11 13/16”), 30cm-42cm(11 13/16” -16 17/32”) 

(Optional), 42cm-48cm(16 17/32

” -18 29/32”) (Optional) 

6.  Weight: approx. 162g (5 23/32oz.) (exclude batteries) 

7.  Measuring method: Oscillometric method, manual inflation and measurement 

8.  Power source: batteries:4 ×1.5V

SIZE AA   

9.  Measurement range: 

Cuff pressure: 0-300mmHg 

Systolic:          60-260mmHg 

Diastolic:        40-199mmHg 

Pulse rate:      40-180 beats/minute 

10.  Accuracy:   

Pressure:        ±3mmHg 

Pulse rate:      ±5% 

11.  Environmental temperature for operation: 5

℃~

40

(41

℉~

104

12.  Environmental humidity for operation: 

90%RH 

13.  Environmental temperature for storage and transport: -20

℃~

55

(-4

℉~

131

14.  Environmental humidity for storage and transport: 

90%RH 

15.  Environmental pressure: 80kPa-105kPa 

16.  Battery life: Approx. 540 times. 

17.  All  components  belonging  to  the  pressure  measuring  system,  including  accessories: 

Pump, Valve, LCD, Cuff and Sensor. 

 

Note: These specifications are subject to change without notice. 

 

NOTICE   
 

1.  Read all of the information in the operation guide and any other literature in the box before 

operating the unit. 

2.  Stay still, calm and rest for 5 minutes before blood pressure measurement. 

3.  The cuff should be placed at the same level as your heart. 

4.  During measurement, neither speak nor move your body and arm. 

5.  Measuring on same arm for each measurement. 

6.  Please always relax at least 1 or 1.5 minutes between measurements to allow the blood 

circulation in your arm to recover.

 

Prolonged over-inflation (

cuff pressure exceed 300 mmHg 

or maintained above15 mmHg for longer than 3 minutes

) of the bladder may cause ecchymoma 

of your arm. 

Summary of Contents for KD-322

Page 1: ...RS 2 INTENDED USE 3 CONTRAINDICATION 3 PRODUCT DESCRIPTION 3 SPECIFICATIONS 4 NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CONNECTING THE CUFF TO THE MONITOR 6 3 APPLYING THE CUFF 6...

Page 2: ...The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peop...

Page 3: ...egrated pressure sensor blood pressure and pulse rate can be measured automatically and non invasively The LCD display will show blood pressure and pulse rate The Electronic Blood Pressure Monitor cor...

Page 4: ...on 90 RH 13 Environmental temperature for storage and transport 20 55 4 131 14 Environmental humidity for storage and transport 90 RH 15 Environmental pressure 80kPa 105kPa 16 Battery life Approx 540...

Page 5: ...ted that you consult with your physician for accurate assessment 13 Please do not use the cuff other than supplied by the manufacturer otherwise it may bring biocompatible hazard and might result in m...

Page 6: ...e 2 CONNECTING THE CUFF TO THE MONITOR Insert the cuff tubing connector into the socket in the left side of the monitor Make certain that the connector is completely inserted to avoid air leakage duri...

Page 7: ...rtably Measurement a Be seated with your feet flat on the floor and don t cross your legs b Place palm upside in front of you on a flat surface such as a desk or table c The middle of the cuff should...

Page 8: ...f the monitor manually h During measurement you can press the exhaust valve which is located in the front of the bulb to release the air Note Please consult a health care professional for interpretati...

Page 9: ...itor 8 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows EE The cuff was not properly applied or the rubber tube was bent or pressed Review the cuff applying and testing sections of...

Page 10: ...iately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a mini...

Page 11: ...use in the electromagnetic environment specified below The customer or the user of the KD 322 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Page 12: ...equency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Gu...

Page 13: ...radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic enviro...

Page 14: ...12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be de...

Reviews: