DN:KD-322-SMSY01 V4.0
Operation Guide
Template version:1
4/14
SPECIFICATIONS
1. Product name: Blood Pressure Monitor
2. Model: KD-322
3. Classification: Internally powered, Type BF applied part,IPX0,No AP or APG,Continuous
operation
4. Machine size: approx. 144mm×102mm×46mm (5 21/32
” x 4 1/32” x 1 13/16”)
5. Cuff circumference: 22cm-30cm(8 21/32
” -11 13/16”), 30cm-42cm(11 13/16” -16 17/32”)
(Optional), 42cm-48cm(16 17/32
” -18 29/32”) (Optional)
6. Weight: approx. 162g (5 23/32oz.) (exclude batteries)
7. Measuring method: Oscillometric method, manual inflation and measurement
8. Power source: batteries:4 ×1.5V
SIZE AA
9. Measurement range:
Cuff pressure: 0-300mmHg
Systolic: 60-260mmHg
Diastolic: 40-199mmHg
Pulse rate: 40-180 beats/minute
10. Accuracy:
Pressure: ±3mmHg
Pulse rate: ±5%
11. Environmental temperature for operation: 5
℃~
40
℃
(41
℉~
104
℉
)
12. Environmental humidity for operation:
≤
90%RH
13. Environmental temperature for storage and transport: -20
℃~
55
℃
(-4
℉~
131
℉
)
14. Environmental humidity for storage and transport:
≤
90%RH
15. Environmental pressure: 80kPa-105kPa
16. Battery life: Approx. 540 times.
17. All components belonging to the pressure measuring system, including accessories:
Pump, Valve, LCD, Cuff and Sensor.
Note: These specifications are subject to change without notice.
NOTICE
1. Read all of the information in the operation guide and any other literature in the box before
operating the unit.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same arm for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood
circulation in your arm to recover.
Prolonged over-inflation (
cuff pressure exceed 300 mmHg
or maintained above15 mmHg for longer than 3 minutes
) of the bladder may cause ecchymoma
of your arm.