Please refer to the cuff circumference range in 

“SPECIFICATIONS” to make 

sure that the appropriate cuff is used.   

2. 

Measuring on same arm each time. 

3. 

Do not move your arm, body, or the monitor and do not move the rubber tube 

during measurement. 

4. 

Stay quiet, calm for 5 minutes before blood pressure measurement. 

5. 

Please keep the cuff clean. If the cuff becomes dirty, remove it from the monitor 

and clear it by hand in a mild detergent, then rinse it thoroughly in cold water. 

Never dry the cuff in clothes dryer or iron it. Clean the cuff after the usage of 

every 200 times is recommended.   

 

4.  BODY POSTURE DURING MEASUREMENT 

 

Sitting Comfortably Measurement 

a.  Be  seated  with  your  feet  flat on  the  floor,  and  don

’t cross 

your legs.   

b.  Place palm upside in front of you on a flat surface such as 

a desk or table 

c. 

The middle of the cuff should be at the level of the right atrium of 

the heart.

 

 

5.  TAKING YOUR BLOOD PRESSURE READING 

 

a.    After  applying  the  cuff  and  your  body  is  in  a  comfortable  position,  press  the 

“ON/OFF”  button.  A  beep  is  heard  and  all  display  characters  are  shown  for 

self-test. 

 

b.  After that

, LCD will show “0” mmHg that indicate you can begin to inflate.     

 

 

c.  Please squeeze the bulb till the pressure is over your normal systolic pressure by 

30mmHg,  if  you  don’t  know  your  normal  systolic  pressure,  please  squeeze  the 

bulb till the pressure to 190mmHg. 

d.  Then  the  monitor  slowly  releases  air  from  the  cuff  and  carries  out  the 

measurement.  Finally  the  blood  pressure  and  pulse  rate  will  be  calculated  and 

displayed on the LCD screen.   

Summary of Contents for KD-322

Page 1: ...RS 2 INTENDED USE 3 CONTRAINDICATION 3 PRODUCT DESCRIPTION 3 SPECIFICATIONS 4 NOTICE 4 SETUP AND OPERATING PROCEDURES 6 1 BATTERY LOADING 6 2 CONNECTING THE CUFF TO THE MONITOR 6 3 APPLYING THE CUFF 6...

Page 2: ...The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most peop...

Page 3: ...egrated pressure sensor blood pressure and pulse rate can be measured automatically and non invasively The LCD display will show blood pressure and pulse rate The Electronic Blood Pressure Monitor cor...

Page 4: ...on 90 RH 13 Environmental temperature for storage and transport 20 55 4 131 14 Environmental humidity for storage and transport 90 RH 15 Environmental pressure 80kPa 105kPa 16 Battery life Approx 540...

Page 5: ...ted that you consult with your physician for accurate assessment 13 Please do not use the cuff other than supplied by the manufacturer otherwise it may bring biocompatible hazard and might result in m...

Page 6: ...e 2 CONNECTING THE CUFF TO THE MONITOR Insert the cuff tubing connector into the socket in the left side of the monitor Make certain that the connector is completely inserted to avoid air leakage duri...

Page 7: ...rtably Measurement a Be seated with your feet flat on the floor and don t cross your legs b Place palm upside in front of you on a flat surface such as a desk or table c The middle of the cuff should...

Page 8: ...f the monitor manually h During measurement you can press the exhaust valve which is located in the front of the bulb to release the air Note Please consult a health care professional for interpretati...

Page 9: ...itor 8 TROUBLESHOOTING 2 PROBLEM POSSIBLE CAUSE SOLUTION LCD Display shows EE The cuff was not properly applied or the rubber tube was bent or pressed Review the cuff applying and testing sections of...

Page 10: ...iately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a mini...

Page 11: ...use in the electromagnetic environment specified below The customer or the user of the KD 322 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environmen...

Page 12: ...equency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Gu...

Page 13: ...radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic enviro...

Page 14: ...12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be de...

Reviews: