ANDILOG CENTOR FIRST II User Manual Download Page 7

 

User's manual 

CENTOR FIRST II 

 

- 7 - 

3.

 

Using the CENTOR FIRST  

  

3.1.

 

 Start-up  

Please note that the Centor First II measures very small changes in force and may not read exactly zero 

if it is moved during its self-test procedure. When properly mounted and zeroed, the reading remains 

stable. 

 

 

 

 

To switch on, press the key   , a short self-test procedure displays the following screen    

  

 

  

   

After the self-test, if the device is functional, the force gauge displays the measurement screen. It then 

displays 0 N. A tare is performed when the device is started. 

 

If the applied force is greater than 20% of the maximum capacity, the  force gauge  displays  a 
maintenance screen. At the bottom of this screen, the force applied in % is indicated. The unit is not 
usable if the force at startup is greater than 20%. It is then necessary to turn off the force gauge and 
remove the preload before turning it on again. 

For example, for a force gauge with a capacity of 500N, the force applied to the sensor must not 
exceed 100N at start-up. 

   

 

Summary of Contents for CENTOR FIRST II

Page 1: ...s manual CENTOR FIRST II User s manual CENTOR FIRST II ANDILOG Technologies BP62001 13845 Vitrolles Cedex France COM TEN Industries 6405 49th Street North Pinellas Park FL 33781 USA Site www andilog...

Page 2: ...4 2 Getting started 6 2 1 Battery charging 6 2 2 Mounting the Accessories 6 2 3 Mounting on a frame 6 3 Using the CENTOR FIRST 7 3 1 Starting up 7 3 2 Extinguishing 8 3 3 Measurement screen 8 3 3 1 D...

Page 3: ...h a battery charge between 30 and 80 of its autonomy It is recommended to use the original power supply provided by Andilog to charge your instrument The power supply must have the following character...

Page 4: ...emperature 20 to 45 C Relative humidity 5 to 95 non condensing Maximum altitude of use 3 000 m 1 2 6 Warranty Subject to the conditions below Andilog Technologies warrants to the purchaser that it wil...

Page 5: ...entation or offers If products are damaged during shipment notify the carrier and Andilog Technologies immediately The warranty is void in case of accident misuse or abuse Calibrations overloaded sens...

Page 6: ...hould be done by hand without applying much torque Excessive torque or the use of pliers can damage the load cell We recommend that you mount the accessories with the force gauge turned on to ensure t...

Page 7: ...evice is functional the force gauge displays the measurement screen It then displays 0 N A tare is performed when the device is started If the applied force is greater than 20 of the maximum capacity...

Page 8: ...n the force gauge is turned off The force gauge will operate with the same settings when it is turned back on unit display order 3 2 Extinguishing To turn off the Centor First II press and hold the bu...

Page 9: ...on will be preceded by the symbol A force applied in compression will be preceded by the symbol The display order can be changed by pressing the 3 3 1 Do the Zero During tests it is often necessary to...

Page 10: ...display The Centor First II remembers the display order when it is turned off 3 3 4 Auto off The automatic shutdown saves the battery charge The force gauge is automatically switched off after 15 min...

Page 11: ...using the on off key My device does not display the measurement screen but an information page with Offset in red If the offset value is greater than 20 the device cannot measure Turn off your device...

Page 12: ...User s manual CENTOR FIRST II 12 Characteristic Capacity Accuracy 0 25 of capacity Autonomy 3 p m Power supply MicroUSB 5V 1A Net weight 520 g Units N daN kg lb Sensor sampling frequency 1000 Hz...

Reviews: