4.
Maintenance and troubleshooting
4.1.
Main faults and solutions
Problem
Solution
My device turns itself off.
1.
Check that the battery percentage is not at 0%.
Put the device on charge
2.
The Centor First II will automatically shut off after
15 minutes if no button is pressed.
My device is blocked, the force does
not vary anymore.
Turn off and restart your instrument using the on/off key.
My device does not display the
measurement screen but an
information page with Offset in red.
If the offset value is greater than 20%, the device cannot
measure. Turn off your device and remove the load
applied to the sensor. Then turn on your device.
If no load is applied to the sensor, then the sensor is
deformed. You must then return your device to us for
replacement of the sensor.
4.2.
Technical specifications
Reference
Capacity
Resolution
CNR FT 10
10 N
0.002 N
CNR FT 25
25 N
0.005 N
CNR FT 50
50N
0.01 N
CNR FT 100
100 N
0.02 N
CNR FT 250
250 N
0.05 N
CNR FT 500
500N
0.1 N