ANDILOG CENTOR EASY Operating Manual Download Page 14

 

Operating Manual – V3.39 

CENTOR EASY Force Gauge  

 

 

- 13 - 

 
 

5.

 

Statistics and memory 

 
The Centor Easy has a mode called “Statistical” allowing to perform a series a measurements and get 
quickly the mean and standard deviation.  
 

a.

 

Limitation 

The memory of the force gauge is limited to 100 measurements; beyond it will be necessary to clear 
the first results to be able to add new ones. 
 

b.

 

Access 

Access to this mode by: pressing “M” button, moving the cursor down to reach the line called 
“STATS” and entering the sub-menu by pressing the right arrow TDX button. 
 

c.

 

Display 

If no action has been carried out on SATS submenu, the screen allows you 
to configure three parameters: 

 

Mes/Samp: Number of measurement per sample 

 

Oper: Personal ID number of the operator 

 

Unit: Measurement unit 

   
If measurements are saved in the memory, those three parameters are 
no longer available. The only way to access it again is by clearing the 
memory. 

 

 
If measurements are saved, they are displayed at the bottom of the page, 
the cursor can move in the table. 
 
The displayed data are as follow: 

 

The first number indicated the sample number, the second 

number is the measure itself 

 

If your samples contains more than one measurement (see 

parameter (Mes/Sample) so you have multiple rows with the 
same sample number 

 
Also 3 statistical calculations are displayed: 

 

Mea: Average of the series of measurements 

 

StDev: standard deviation of the series of measurements 

 

S/M: Standard deviation divided by the average 

 
 

d.

 

Possible actions 

 
If no data are saved, you can configure the three parameters described above. To do so place the 
cursor on the parameter and change the settings with the left and right button 
 
If data are saved, you can: 

 

Review the result, moving the cursor in the table: up and down keys 

 

Delete the selected sample or the entire series: left button (ZERO) 

 

Send data : right button (TDX) 

M e s / S a m p       1  
O p e r .                 0 1  
U n i t                       N  
 
 
 
 

M e s / S a m p       1  
O p e r .                 0 1  
U n i t                       N  
 
S T A T S  
S a m p  
M e a  
S t D e v  
S / M  
 
0 0 1     + x . x x x  
0 0 2     + x . x x x  
0 0 3     + x . x x x  
 

Summary of Contents for CENTOR EASY

Page 1: ...ng Manual CENTOR EASY Gauge Operating Manual CENTOR EASY V3 39 ANDILOG Technologies BP62001 13845 Vitrolles Cedex 9 France E mail info andilog com Web www andilog com Ph 33 0 442 348 340 Fax 33 0 442...

Page 2: ...sure unit 7 6 Peak values 7 7 Backlit 7 V Advanced features 8 1 Limits 9 2 Display 10 3 IN OUT 11 4 RS232 12 5 Statistics and memory 13 6 System 15 VI Associated products 16 1 Manual or motorised test...

Page 3: ...that the gauge has not been damaged Unpacking The CENTOR EASY force gauge is supplied with Carrying case Extension rod Internal sensor only Hook Internal sensor only Compression plate Internal sensor...

Page 4: ...ltitude 3000m 4 Battery The battery life is 8 hours under a normal use The gauge should be charged after normal use You should charge it every 3 weeks to ensure a maximum life time of the batteries It...

Page 5: ...ge 4 IV Presentation of your force gauge 1 Front panel scheme CAPACITY DATA TRANSFER CHANGES UNIT CHANGES DISPLAYED VALUE ZEROING ON OFF SETTING MENU BACKLIGHT CANCEL ACTION INTERNAL SENSOR AREA MAIN...

Page 6: ...tton the information screen showing the status of the CENTOR force gauge is displayed during 5 seconds and then the main screen is displayed CONNECTOR 9 PINS OFF ON CENTOR EASY CONNECTOR 15 PINS INCLU...

Page 7: ...tion the weight of accessories hook or plate fixed on the sensor s rod Pressing this button reset all memories especially peak values NOTA When turned on the CENTOR achieves several tests especially t...

Page 8: ...imum and systematically displays them on line 2 Pressing MAX button makes max value min value and current value successively display as well as a calculation if it has been set 7 Backlit You can turn...

Page 9: ...menu uses MAX and UNIT buttons To select the required function highlights it and press the TDX button To exit uses M button Save and return to previous screen Move the cursor in DOWN direction Move th...

Page 10: ...e or deactivate this feature Once the feature is activated the parameters are displayed In the main measuring screen the status line is modified signs are displayed they display if the value is inside...

Page 11: ...a motorized test stand Note If only one threshold is necessary max value for example enter the max or min capacity of the sensor for the second threshold value 2 Display Exit by pressing the M button...

Page 12: ...mpulsion I negative impulsion I Change to positive state M change to negative state M Open O Closed F Anal INT NO Active or de active the analog output Digi NO VLI MXI MNI Active the Digimatic output...

Page 13: ...Number of bits Stop 1 2 Number of stop bits The second part defines the characters sequence sent Menu Choice Description Demand F Current Value P maximum V minimum C Inactive T Last 100 values in memo...

Page 14: ...g the memory If measurements are saved they are displayed at the bottom of the page the cursor can move in the table The displayed data are as follow The first number indicated the sample number the s...

Page 15: ...any time the maximum is displayed in large characters on line 1 You can if you wish display the current value instead of the maximum by pressing the top button MAX When a measurement was made press ri...

Page 16: ...39 CENTOR EASY Force Gauge 15 6 System The following settings can be set Date Hour and Language Version of the internal force gauge Date of the day Language Serial number Hour of he day Due date of t...

Page 17: ...ping accessories adapted to different tests hooks plates pliers self squeezing jaws eccentric jaws peeling fixture etc If you want to receive our catalogue contact your ANDILOG TECHNOLOGIES dealer 3 A...

Page 18: ...ration and reload it at any time To do so press the M key to go to the Menu mode then Press ZERO and M to save the configuration Press TDX and M to reload the saved configuration Press X and M to load...

Page 19: ...ettings Feature Value LIMITS Chanel NO DISPLAY LINE1 VLI LINE2 MXI SIGN TRA DECIMA 4 BARG YES RAFF N DIRECT H AUTOFF 10 IN OUT PEDALE RAZ TOP1 F ANAL NO DIGI NO RS232 NO RS232 Bds 9600 Par None Bits 8...

Page 20: ...15 points of CENTOR electronic card PIN 1 Not connected 2 Reset To be connected to Ground for reset 3 RS232 TXD 4 RS232 RXD 5 Digimatic 6 Digimatic 7 Digimatic 8 Digimatic 9 Output Force N 1 10 Output...

Page 21: ...Operating Manual V3 39 CENTOR EASY Force Gauge 20 6 Dimensions...

Reviews: