![AND TM-2440 Instruction Manual Download Page 91](http://html.mh-extra.com/html/and/tm-2440/tm-2440_instruction-manual_2941339091.webp)
73
Immunity test
IEC60601-1-2
test level
Compliance
level
Electromagnetic
environment
Radiated RF
IEC61000-4-3
3
V/m
80
MHz to
2.5
GHz
10
V/m
80
MHz to
2.6
GHz
Recommended
separation distance :
d
=
1.2
80
MHz to
800
MHz
d
=
2.3
800
MHz to
2.5
GHz
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
<
5
%
U
T
(
>
95
% dip in
U
T
)
for
0.5
cycle
<
40
%
U
T
(
>
60
% dip in
U
T
)
for
5
cycle
<
70
%
U
T
(
>
30
% dip in
U
T
)
for
25
cycle
<
5
%
U
T
(
>
95
% dip in
U
T
)
for
5
seconds
N.A.
Does not apply
because the power
supply unit is built in.
Note :
U
T
is the AC mains voltage prior to application of the test level.