Electromagnetic immunity
Immunity test
IEC60601-1-2
test level
Compliance
level
Electromagnetic
environment
Electrostatic
discharge
(ESD)
IEC61000-4-2
±
6
kV
contact
±
8
kV
in air
±
8
kV
contact
±
15
kV
in air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±
2
kV for power
supply lines
±
1
kV for
input/output lines
N.A.
Surge
IEC61000-4-5
±
1
kV differential
mode
±
2
kV common
mode
N.A.
Does not apply
because the built-in
power supply unit is
built-in.
Power
frequency
(
50
/
60
Hz)
magnetic field
IEC61000-4-8
3
A/m
30
A/m
Power frequency
magnetic fields should
become at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Conducted RF
IEC61000-4-6
3
Vrms
150
kHz to
80
MHz
6
Vrms
150
kHz to
80
MHz
Recommended
separation distance :
d
=
1.2