7. Sensitivity Adjustment / Calibration Test
7.1. Automatic sensitivity adjustment
The sensitivity of the balance is automatically adjusted using the internal mass according to ambient
temperature change, set time, or interval time. For details, refer to the separate instruction manual listed
in “1.1. Detailed Manuals”.
7.2. Sensitivity adjustment using the internal mass
The sensitivity adjustment / calibration test screen is displayed with the
button on the weighing
screen.
Name
Description
1
External sensitivity
adjustment
Your external calibration weight is used to adjust the balance.
Instructions are displayed on the screen for how to adjust the
sensitivity.
2
Internal sensitivity
adjustment
The internal mass is used to automatically adjust the balance.
3
Internal calibration test
The internal mass is used to check the accuracy of weighing, and
the result is output.
Note that the balance is not adjusted.
4
External calibration test
Your external calibration weight is used to check the accuracy of
weighing, and the result is output.
Note that the balance is not adjusted.
Do not allow vibration or drafts to affect the balance during sensitivity adjustment / calibration test.
For details, refer to the separate instruction manual listed in “1.1. Detailed Manuals”.
2. Internal
sensitivity
adjustment
3. Internal
calibration
test
1. External
sensitivity
adjustment
4. External
calibration
test