Maintenance and care
UV lamp for InnuPure C16
Edition 03/2013
17
7
Maintenance and care
The UV lamp is largely maintenance-free. Care and maintenance work is restricted to clean-
ing/disinfecting the device and changing the fuse.
CAUTION!
The device cover must only be opened by the technical customer service of Analytik Jena
AG or its authorized agency. Before opening the device cover, disconnect the mains plug!
When an exchange of the illuminant is necessary, please contact the Analytik Jena service
or its representatives.
7.1
Cleaning the UV lamp for the InnuPure C16
CAUTION! Danger of electric shock!
Before cleaning the device with disinfection agents, disconnect the mains plug from the
mains socket!
The UV lamp must only be recommissioned once it is completely dry!
Clean the UV lamp as follows:
If the InnuPure C16 is used for the analysis of infectious material, great care must
be taken. Contaminations may be transferred to the UV lamp via the guide rails.
Visible contamination inside the InnuPure C16 must be removed immediately using
suitable detergents, making sure that no solvent enters the inside of the device.
The contact areas between UV lamp and interior of the InnuPure C16 must be
cleaned and disinfected with appropriate disinfection cloths.
As a possible disinfection agent we recommend
Meliseptol HBV – cloths B. Braun company
If the UV lamp of an InnuPure C16 has to be sent to Analytik Jena AG for service
after infectious material has been processed with it, a prior decontamination must
be carried out and documented.