Site requirements
UV lamp for InnuPure C16
Edition 03/2013
13
5
Site requirements
5.1
Installation conditions
The following requirements apply to the climatic conditions in the operating room of the UV
lamp (corresponding to those of the InnuPure C16):
Temperature range:
+5 °C to +35 °C
Max. humidity:
80 % at 30 °C
5.2
Space requirement
The space requirement results from the device dimensions:
W x H x D
240 mm x 120 mm x 410 mm
The UV lamp is moved through the opening in the door of the InnuPure C16. The UV lamp
can slid into the lower level of the priming station of the InnuPure C16 for storage purposes.
CAUTION!
Before sliding the UV lamp into the InnuPure C16, the tray has to be removed from the ex-
traction device and the InnuPure C16 must be switched off!
Due to danger of tipping, the storage of the UV lamp on the upper level of the priming station
of the InnuPure C16 is only possible when the upper rear stopper plate is removed! After
removing the rear stopper, the front stopper can be slid into the priming station.
5.3
Energy supply
WARNING!
The UV lamp for the InnuPure C16 must only be connected to a properly grounded mains
outlet in accordance with the voltage specifications on the type plate!
The UV lamp for the InnuPure C16 is operated on single-phase alternating current. The in-
stallation of the electrical equipment of the laboratory must comply with the standard DIN
VDE 0100. After the connection point, an electrical current in accordance with the standard
IEC 38 must be available.