Startup/operation
14
Edition 03/2013
UV lamp for InnuPure C16
6
Startup/operation
CAUTION!
The InnuPure C16 must be switched off before the UV lamp is inserted and the door of the
InnuPure C16 must be closed. The UV lamp must be completely slid into the InnuPure C16
before the mains plug may be connected to the mains socket.
CAUTION! Danger of short-circuit from condensate!
If the storage and the installation temperature differ a lot, wait until the UV lamp has adapted
to the new ambient temperature before connecting it in order to prevent damages to the de-
vice by condensation water.
IMPORTANT
Retain the transport packaging! Return transport for service must be in the original packag-
ing. This alone prevents transport damage.
6.1
Switching on the UV lamp for the InnuPure C16
1. Remove the tray from the InnuPure C16 and switch off the InnuPure C16.
2. Set the desired operating time via the time relay (see chapter "Setting the operating
time of the UV lamp with the time relay" p. 15).
3. Make sure the door of the InnuPure C16 is closed before inserting the UV lamp. Com-
pletely slide the UV lamp into the InnuPure C16 until the stop of the rear of the front
plate.
4. Connect the mains plug to the UV lamp and to the mains socket.
5. Switch on the UV lamp via the front button and check whether the UV lamp gives light.
Note:
Monitor the settings of the operating time via the time relay!
The status LED is lit green during the set operating time.