Page 1: ...UV lamp Decontamination set for InnuPure C16 Operating manual ...

Page 2: ...ation about Analytik Jena AG on the internet http www analytik jena de Documentation number 10 3221 010 23 Edition March 2013 Implementation of the Technical Documentation Analytik Jena AG This publication describes the state of this product at the time of publishing It need not necessarily agree with future versions of the product Modifications reserved Copyright 2013 Analytik Jena AG ...

Page 3: ...l resistance of the device 9 3 9 Behavior during emergencies 10 3 10 Standards and directives 10 4 Technical description 11 4 1 Fundamental layout and function 11 4 2 Design connections indications 11 5 Site requirements 13 5 1 Installation conditions 13 5 2 Space requirement 13 5 3 Energy supply 13 6 Startup operation 14 6 1 Switching on the UV lamp for the InnuPure C16 14 6 2 Switching off the U...

Page 4: ...g 3 Total view of the UV lamp for the InnuPure C16 12 Fig 4 Time relay on the right side of the device 12 Fig 5 Safety switch on the left side of the device 12 Fig 6 Inserting lamp into InnuPure C16 connecting and switch on 15 Fig 7 Time relay on the right side of the device 16 Fig 8 Operating time of the UV lamp at different settings 16 Fig 9 Open fuse holder 18 ...

Page 5: ...of the picto graphs and signal words used is explained in the chapter Safety notes 1 2 Intended use The UV lamp is an optional accessory for the InnuPure C16 and is used for the internal con tamination The InnuPure C16 is a flexible and efficient extraction system for the fully auto mated isolation and purification of nucleic acid in chemical and biological laboratories In the field of medicine an...

Page 6: ...re excluded for personal injury and property damage due to one or several of the following causes usage of the UV lamp for the InnuPure C16 not according to the intended purpose improper commissioning operation and service of the UV lamp modifications of the UV lamp without prior consultation with Analytik Jena AG unauthorized handling inside the UV lamp operation of the UV lamp with faulty safety...

Page 7: ...g Little contact surfaces with the interior of the InnuPure C16 guide rails easy to disinfect Other technical data Weight Approx 2 5 kg Dimensions W x H x D 240 mm x 120 mm x 410 mm Electrical variables Power supply 230 V AC 50 60 Hz Protection 1 x 160 mA T 5x20 mm 250V max power consumption 30 W Environmental conditions Temperature during op eration 5 35 C Humidity during opera tion Max 80 at 30 ...

Page 8: ...rotection have to be observed and complied with References to potential dangers do not replace the work protection regulations which must be observed 3 2 Symbols and signal words used in the user manual The user manual uses the following symbols and signal words to indicate hazards or instruc tions The safety instructions are always placed before an action CAUTION Indicates a potentially hazardous...

Page 9: ...aning Remark Warning against a hazard location Warning of UV radia tion Warning of UV radiation in tended use only inside the InnuPure C16 with closed door Notice symbol Meaning Remark Before opening the device always disconnect the mains plug Before opening the device Switch off the device and disconnect the mains plug from the mains socket and the device Observe the instructions for use Prior to...

Page 10: ...r viced by trained personnel instructed in technical safety The operation or servicing of the UV lamp by minors or individuals under the influ ence of alcohol drugs or medication is not permitted It must be ensured that only authorized personnel works at the UV lamp The operating personnel must be familiar with the dangers arising from the sub stances used inside the InnuPure C16 The appropriate p...

Page 11: ... prevent fluids from entering the UV lamp through openings or gaps Do not short circuit the UV lamp and only use the fuse in accordance with the specifications in the technical data 3 8 Handling hazardous substances Even with intended use there is a risk of health damage when handling hazardous substanc es The operator is solely responsible for the compliance with all safety requirements to pro te...

Page 12: ...hers must be within reach and easy to ac cess All equipment has to be in a sound condition and should be checked regular ly 3 11 Standards and directives The UV lamp was manufactured according to the currently valid technology regulations and the approved safety related regulations When constructing the UV lamp the relevant safety and health requirements of the applica ble laws standards and regul...

Page 13: ...effective for the destruction of bacteria fungi yeasts and viruses In addition the UV lamp denatures unwanted DNA and RNA inside the InnuPure C16 Fig 2 UV C radiation killing microorganisms 1 UV Technik Meyer GmbH Datenblatt Desinfektion mit UV C D v1 6 2 AR 28 10 2009 4 2 Design connections indications Beside the mains connection the device front provides the button to switch on the UV lamp Durin...

Page 14: ...nnuPure C16 1 Time relay 2 Circuit closer 3 Status LED of the UV lamp 4 Mains connection 5 Fuse holder Fig 3 Total view of the UV lamp for the InnuPure C16 Fig 4 Time relay on the right side of the device Fig 5 Safety switch on the left side of the device ...

Page 15: ...uPure C16 the tray has to be removed from the ex traction device and the InnuPure C16 must be switched off Due to danger of tipping the storage of the UV lamp on the upper level of the priming station of the InnuPure C16 is only possible when the upper rear stopper plate is removed After removing the rear stopper the front stopper can be slid into the priming station 5 3 Energy supply WARNING The ...

Page 16: ...packaging Return transport for service must be in the original packag ing This alone prevents transport damage 6 1 Switching on the UV lamp for the InnuPure C16 1 Remove the tray from the InnuPure C16 and switch off the InnuPure C16 2 Set the desired operating time via the time relay see chapter Setting the operating time of the UV lamp with the time relay p 15 3 Make sure the door of the InnuPure...

Page 17: ...om the UV lamp and the mains socket 3 Remove the UV lamp from the InnuPure C16 CAUTION Warming of the housing in the illumination area Depending on the operating time the housing of the UV lamp may be heated up to 60 in the illumination area 6 3 Setting the operating time of the UV lamp with the time relay Depending on the coding of the DIP switches 1 2 3 the operating time of the UV lamp can be s...

Page 18: ... side of the device Operating time in dependence of coding setting value Setting value of Time relay DIP switch coding 0 5 minimum 0 05s 0 5s 5s 0 5min 5min 0 5h 5h 1 0 1s 1s 10s 1min 10min 1h 10h 5 0 5s 5s 50s 5min 50min 5h 50h 10 maximum 1s 10s 100s 10min 100min 10h 100h Fig 8 Operating time of the UV lamp at different settings ...

Page 19: ...he mains plug from the mains socket The UV lamp must only be recommissioned once it is completely dry Clean the UV lamp as follows If the InnuPure C16 is used for the analysis of infectious material great care must be taken Contaminations may be transferred to the UV lamp via the guide rails Visible contamination inside the InnuPure C16 must be removed immediately using suitable detergents making ...

Page 20: ... The fuse holder comprises an addi tional fuse for the first exchange 1 Disconnect the mains plug from the mains socket 2 Open the fuse holder by pulling its cover 3 Replace the faulty mains fuse rear fuse Use the following fuse 1 x T 160 mA 250 V 4 Close the fuse holder 5 Dismantle the UV lamp as described in section Startup operation p 14 1 Equipment fuse 2 Replacement fuse 3 Fuse holder Fig 9 O...

Page 21: ...rted in such a way that major temperature fluctuations are avoided and the formation of condensate is thus prevented 8 2 Storage ATTENTION Environmental influences and condensate formation can destroy individual components of the UV lamp The storage atmosphere of the UV lamp must be low in dust and free from aggressive va pors If the device is not installed immediately after delivery or not requir...

Page 22: ...for InnuPure C16 9 Disposal ENVIRONMENTAL PROTECTION At the end of its service life the UV lamp for the InnuPure C16 and all its electronic compo nents must be disposed of in accordance with the applicable regulations as electronic waste ...