Safety instructions
Biometra TOne
8
Version 01.21
The use of oil between the samples and the sample block is not necessary to achieve
an improved heat exchange. However, if you still want to use oil, you should use
mineral oil. Do not use silicone oil.
Be careful to avoid crushing or pinching injuries when closing the device.
Biometra TOne may cause burns. The following has to be observed:
The thermal block, the samples, and the heated lid reach high temperatures. There is
a risk of burns during contact.
Ensure that the lid is securely closed before starting the program! Do not touch the
heated lid!
The rapid heating of the thermal block can cause liquids to boil explosively. Always
wear safety goggles during operation!
Do not touch hot sample tubes or plates and do not open them or boiling liquid may
escape!
Only use plates and tubes that are suitable for high temperatures (up to 100 °C), fit
well into the thermal block (no wobbling), and whose lids seal tightly!
Modifications and extensions to the device may only be carried out after consultation
with Analytik Jena. Unauthorized changes can limit the operating safety and lead to
restrictions in warranty and access to customer service.
2.6.2
Safety instructions: protection against explosion and fire
Biometra TOne must not be operated in an explosive environment. Biometra TOne must
not be operated with flammable, explosive, or volatile substances.
2.6.3
Safety instructions: electrical equipment
Work on the electrical components of Biometra TOne may only be performed by a quali-
fied electrical technician in accordance with the applicable electrical engineering rules.
Life-threatening electrical voltages may occur in the interior of Biometra TOne!
The following has to be observed:
Any work on the interior of the device may only be carried out by the Analytik Jena
service department and specially authorized technicians.
The electrical components must be checked regularly by a qualified electrician. Any
defects, such as loose connections, faulty, or damaged cables, must be repaired with-
out delay.
Before opening the device it must be switched off at the main power switch and the
power plug must be disconnected from the power outlet!
Biometra TOne must be switched off immediately using the main power switch and
the power plug disconnected from the power supply if any faults occur in the electric
components.
Do not operate the device near sources of strong electromagnetic radiation (e.g., a
non-shielded, deliberately operated high frequency source), as these may affect
proper operation of the device.